ENDOPATH
Report
- Report Number
- 3005075853-2017-05787
- Event Type
- Injury
- Date Received
- October 30, 2017
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K020779
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). WE DID NOT RECEIVE A BATCH OR LOT NUMBER FOR THE PRODUCT INVOLVED IN THIS COMPLAINT. THEREFORE, WE WERE UNABLE TO CHECK MANUFACTURING RECORDS FOR ANY RELATED NON-CONFORMANCE.
IT WAS REPORTED THAT DURING REVIEW OF JOURNAL ARTICLE: DANGERS IN THE USE OF STAPLERS IN LIVER SURGERY. AUTHORS: ANGELA RIGA, NARIMAN KARANJIA. CITATION: ANN R COLL SURG ENGL 2007; 89: 810¿811. THE AUTHORS REPORT 2 CASES WHERE THE ENDOPATH ETS FLEX 45 ENDOSCOPIC ARTICULATING LINEAR CUTTER FAILED WITH DANGEROUS CONSEQUENCES. THE FIRST CASE WAS A (B)(6) WOMAN WITH A COLORECTAL METASTASIS STRADDLING SEGMENTS 2 AND 4A UNDERWENT A LEFT HEMIHEPATECTOMY. AFTER DIVISION OF THE HILAR STRUCTURES AND COMPLETION OF THE PARENCHYMAL TRANSECTION, THE LEFT MIDDLE VEIN COMPLEX WAS STAPLED WITH ENDOPATH ETS FLEX 45 ENDOSCOPIC ARTICULATING LINEAR CUTTER. THE STAPLER FIRED STAPLES WHICH DID NOT CLOSE AND THE VEIN WAS DIVIDED WITH CATASTROPHIC BLEEDING. THE HAEMORRHAGE WAS CONTROLLED WITH DIGITAL PRESSURE AND THE OPENING IN THE VENA CAVA UNDER-RUN WITH PROLENE. THE BLOOD LOSS WAS 1000ML, BUT TRANSFUSION WAS AVOIDED AND SHE MADE A FULL RECOVERY. THE 2ND CASE INVOLVED A (B)(6) MAN WITH A COLORECTAL METASTASIS STRADDLING SEGMENTS 4A, 8 AND 7 UNDERWENT A RIGHT HEMI-HEPATECTOMY EXTENDED TO SEGMENT 4A. AFTER EXPOSING AND TAPING THE RIGHT ANTERIOR AND POSTERIOR SECTION TRUNKS, THE ANTERIOR SECTION TRUNK WAS STAPLED WITH ENDOPATH ETS FLEX 45 ENDOSCOPIC ARTICULATING LINEAR CUTTER. THE STAPLER DIVIDED THE TRUNK WITHOUT A ROW OF STAPLES ON THE PROXIMAL END LEADING TO PROFUSE BLEEDING. THE ENSUING HAEMORRHAGE WAS CONTROLLED WITH PRESSURE AND THE ANTERIOR SECTION TRUNK WAS OVERSEWN WITH A PROLENE WHIP STITCH. THE TOTAL OPERATIVE BLOOD LOSS IN THIS PATIENT WAS 2L AND HE REQUIRED A 4-UNIT BLOOD TRANSFUSION, BUT HE MADE A FULL RECOVERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 765607 | ENDOPATH | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |