MAYFIELD INFINITY XR2 BASE UNIT
Report
- Report Number
- 3004608878-2017-00305
- Event Type
- Malfunction
- Date Received
- October 30, 2017
- Date of Event
- October 17, 2017
- Report Date
- October 19, 2017
- Manufacturer
- INTEGRA LIFESCIENCES CORPORATION OH/USA
- Product Code
- FWZ
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WAS RECEIVED ON 08NOV2017 AS FOLLOWS: ON (B)(6) 2017, A (B)(6) MALE PATIENT WAS TO UNDERGO AN OSTEOTOMY OF THE MIDDLE FOSSA WITH DURAL TUMOR EXCISION. THE REPORTED EVENT DID NOT LEAD TO INCREASE OF THE SURGERY TIME. THE MEDICAL STAFF FINISHED THE PROCEDURE USING ANOTHER POSITIONING SYSTEM. INTEGRA HAS COMPLETED THEIR INTERNAL INVESTIGATION ON (B)(6) 2017. THE REPORTED ERROR BY THE CUSTOMER WAS CONFIRMED. THE PART S437A2309 CLAMP KNOB ASSEMBLY WAS BROKEN. DEVICE HISTORY REVIEW: CAN NOT BE PERFORMED AT THIS TIME AS NO SERIAL NUMBER OR LOT# HAVE BEEN PROVIDED. CAPA HAS BEEN ISSUED TO FURTHER INVESTIGATE THIS REPORTED FAILURE AND AS SUCH A TREND ANALYSIS WILL BE CAPTURED IN THIS PROJECT. INTEGRA LIFE SCIENCES THOROUGHLY INVESTIGATED THE COMPLAINT EVENT. THIS INVESTIGATION INCLUDED OUTSOURCED MATERIAL TESTING, STRESS ENGINEERING AND FRACTOGRAPHY. THE CONSENSUS OF INVESTIGATION FINDS THE MOST LIKELY ROOT CAUSE OF COMPLAINT EVENT IS: USER APPLIED TORQUE TO KNOB IN EXCESS OF MATERIAL STRENGTH OF SCREW THREAD OR COMBINATION OF USER TORQUE AND CLAMP SUPPORT LOAD EXCEEDING SCREW THREAD STRENGTH.
THE SCREW OF THE (B)(4) STANDARD KNOB ASSEMBLY (PART OF THE MAYFIELD INFINITY XR2 BASE UNIT A2079) BROKE DURING USE, WHEN THE FINAL USER WAS POSITIONING THE PATIENT. THE PATIENT WAS ANESTHETIZED AND THERE WAS NO PATIENT INJURY. THE SURGICAL PROCEDURE WAS PERFORMED. REQUEST FOR ADDITIONAL INFORMATION HAS BEEN SENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 766344 | MAYFIELD INFINITY XR2 BASE UNIT | MAYFIELD | FWZ | INTEGRA LIFESCIENCES CORPORATION OH/USA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |