FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 6984493 · Received October 27, 2017

Report

Report Number
3008642652-2017-08992
Event Type
Malfunction
Date Received
October 27, 2017
Date of Event
September 27, 2017
Report Date
October 27, 2017
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (MONITOR RESETS) WAS CONFIRMED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO A DEFECTIVE U104 PXA PROCESSOR ON THE COMPUTER/ANALOG BOARD. THE ROOT CAUSE FOR THE DEFECTIVE U104 PXA PROCESSOR COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR.

Description of Event or Problem · 1

A US DISTRIBUTOR REPORTED THAT A MONITOR WAS RESETTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764582 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 00855778005005

Patients

Seq Age Sex Outcome Treatment
1