Narratives
TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
INTEGRA HAS COMPLETED THEIR INTERNAL INVESTIGATION ON NOVEMBER 9, 2017. RESULTS: DHR REVIEW; THE MANUFACTURING RECORDS FOR PRODUCT ID TRL-0920-025-14 COULD NOT BE REVIEWED DUE TO THE LOT NUMBER NOT PROVIDED BY THE COMPLAINANT FOR THE REPORTED INCIDENT ALLEGED. COMPLAINTS HISTORY; A REVIEW OF COMPLAINT RECORDS DURING THE LAST 5 YEARS FOUND 16 HUMERAL STEM TRIAL COMPLAINTS REPORTED FOR BREAKAGE. DURING THAT PERIOD OF TIME, THERE HAVE BEEN APPROXIMATELY 3529 TSS SURGERIES REPORTED. THIS REPRESENTS A COMPLAINT RATE OF 0.45% (16/3529) AS SOME OF THE REPORTED EVENTS WERE ASSOCIATED WITH A SERIOUS SEVERITY LEVEL, THIS INDICATES AN ADVERSE TREND. CONCLUSION: AT THIS TIME, THE ROOT CAUSES FOR THIS INCIDENT COULD NOT BE DETERMINED WITH THE AVAILABLE INFORMATION.
THE 1 OF 2 REPORTS: SAME PATIENT, SAME PROCEDURE, DIFFERENT PRODUCTS. OTHER MFG REPORT #:1651501-2017-00041. IT WAS REPORTED THAT DURING THE PREPARATION OF THE HUMERAL CANAL FOR A REVERSE SHOULDER ARTHROPLASTY, THE HUMERAL STEM TRIAL BROKE OFF THE STEM TRIAL HANDLE. THIS CASE WAS A REVISION OF A PRIMARY FROM THE PREVIOUS WEEK THAT HAD DISLOCATED. THE IMPLANTED STEM WAS A SIZE 13, BUT THE SURGEON SIZED UP TO A 14 FOR THE REVERSE TO INCREASE THE BODY HEIGHT. THE SURGEON HAD TO MAKE ANOSTEOTOMY TO BE ABLE TO REMOVE THE IMPACTED STEM TRIAL. THE EVENT LEAD TO 90 MINUTES SURGICAL DELAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 762483 | HUMERAL STEM TRIAL, 14MM | TITAN TOTAL SHOULDER INSTRUMENTS | HWT | ASCENSION ORTHOPEDICS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |