FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 6982707 · Received October 27, 2017

Report

Report Number
3008382007-2017-22975
Event Type
Malfunction
Date Received
October 27, 2017
Report Date
October 10, 2017
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K082590
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6), THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE ERRATIC RESULTS ON THE SUBJECT METER OF "59, HI, 240, 176, 92, 178 AND 103 MG/DL", PERFORMED WITHIN 20 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762486 OT PING METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3988297

Patients

Seq Age Sex Outcome Treatment
1 112 YR