FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM

MDR report key: 6980538 · Received October 26, 2017

Report

Report Number
1723170-2017-04331
Event Type
Malfunction
Date Received
October 26, 2017
Date of Event
September 29, 2017
Report Date
January 10, 2018
Manufacturer
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
Product Code
HAW
UDI-DI
00613994653482
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. A MEDTRONIC REPRESENTATIVE WENT TO SITE TO TEST THE EQUIPMENT. REPRESENTATIVE WAS UNABLE TO REPLICATE THE REPORTED ISSUE. A SYSTEM CHECKOUT WAS PERFORMED AND THE HARDWARE, SOFTWARE, AND INSTRUMENTS PASSED THE SYSTEM CHECKOUT. THE SYSTEM WAS FOUND TO BE FULLY FUNCTIONAL. NO PARTS WERE REPLACED. NO PARTS HAVE BEEN RECEIVED BY MANUFACTURER FOR EVALUATION.

Additional Manufacturer Narrative · 1

THE LOGS FOR THE NAVIGATION SYSTEM WERE REVIEWED BY MEDTRONIC PERSONNEL. HOWEVER, THE LOGS PROVIDED NO ADDITIONAL INSIGHT INTO THE PROBABLE CAUSE OF THE ANOMALY. THE BEHAVIOR OF THE REPORTED ISSUE WAS FOUND TO BE CONSISTENT WITH A KNOWN ANOMALY IN THE MEDTRONIC NAVIGATION SOFTWARE ANOMALY TRACKING DATABASE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION SOFTWARE ANOMALY TRACKING DATABASE.

Description of Event or Problem · 1

A SITE REPRESENTATIVE REPORTED THAT WHILE OUTSIDE OF PROCEDURE, WHEN ON "SELECT PROCEDURE" SCREEN THE NAVIGATION SYSTEM BECAME UNRESPONSIVE. SITE HARD SHUTDOWN THE RACK AND NAVIGATION INTERFACE UNIT (NIU). AFTER WAITING FOR ABOUT A MINUTE THE NAVIGATION INTERFACE UNIT (NIU) WAS TURNED ON AND THEN THE RACK WAS TURNED ON. UPON BOOTING UP THE SYSTEM FUNCTIONED NORMALLY. THERE WAS NO PATIENT PRESENT AT THE TIME OF THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
760465 STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. (LOUISVILLE) I7 00613994653482

Patients

Seq Age Sex Outcome Treatment
1