Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) REPLACED POWER MANAGEMENT (PM) BOARD, MOTOR CONTROLLER (MC) BOARD AND DATA ACQUISITION (DA) BOARD TO ADDRESS THE REPORTED PROBLEM.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS ON THE RETURNED POWER MANAGEMENT (PM) BOARD, DATA ACQUISITION (DA) BOARD AND MOTOR CONTROLLER (MC) BOARD SHOWS THAT THE U28 PIN 6 INTERNAL SHORTED TO PIN 8 (GND) ON THE MOTOR CONTROLLER PCBA AND THE U16 PIN 6 INTERNAL SHORTED TO PIN 8 (GND) ON THE POWER MANAGEMENT PCBA CREATED THE 1007 ERROR CODE VENT INOP.

Additional Manufacturer Narrative · 1

A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

THE FIELD SERVICE ENGINEER (FSE) STATES THAT THE SERVICE SCREENS SHOWS XXXXXX AND NO OTHER INFO FOR EVERY BOARD EXCEPT THE CPU BOARD. T SHUT DOWN..THE F.S.E. REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
755284 V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE MNT RESPIRONICS CALIFORNIA, INC V60

Patients

Seq Age Sex Outcome Treatment
1