V60 VENTILATOR
Report
- Report Number
- 2031642-2017-03420
- Event Type
- Malfunction
- Date Received
- October 25, 2017
- Report Date
- October 4, 2017
- Manufacturer
- RESPIRONICS CALIFORNIA, INC
- Product Code
- MNT
- PMA / PMN Number
- K082660
- Removal / Correction Number
- Z-2061-2017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THE FIELD SERVICE ENGINEER (FSE) REPLACED POWER MANAGEMENT (PM) BOARD, MOTOR CONTROLLER (MC) BOARD AND DATA ACQUISITION (DA) BOARD TO ADDRESS THE REPORTED PROBLEM.
FAILURE ANALYSIS ON THE RETURNED POWER MANAGEMENT (PM) BOARD, DATA ACQUISITION (DA) BOARD AND MOTOR CONTROLLER (MC) BOARD SHOWS THAT THE U28 PIN 6 INTERNAL SHORTED TO PIN 8 (GND) ON THE MOTOR CONTROLLER PCBA AND THE U16 PIN 6 INTERNAL SHORTED TO PIN 8 (GND) ON THE POWER MANAGEMENT PCBA CREATED THE 1007 ERROR CODE VENT INOP.
A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.
THE FIELD SERVICE ENGINEER (FSE) STATES THAT THE SERVICE SCREENS SHOWS XXXXXX AND NO OTHER INFO FOR EVERY BOARD EXCEPT THE CPU BOARD. T SHUT DOWN..THE F.S.E. REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 755284 | V60 VENTILATOR | VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE | MNT | RESPIRONICS CALIFORNIA, INC | V60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |