FDA Adverse Event Malfunction Summary report: N

20ML DISCARDIT II SYRINGE W/O NEEDLE

MDR report key: 6977590 · Received October 25, 2017

Report

Report Number
3002682307-2017-00072
Event Type
Malfunction
Date Received
October 25, 2017
Date of Event
September 29, 2017
Report Date
November 16, 2017
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SAMPLE EVALUATION: WE HAVE BEEN PROVIDED WITH THE AFFECTED SAMPLES. AFTER THE EVALUATION OF THE RETURNED SAMPLES, WE CONFIRMED THE REPORTED ISSUE, AND CONCLUDED THAT THE WHITE PARTICLES INSIDE THE SYRINGE WERE COMPOSED BY LUBRICANT FROM THE SYRINGE BARREL. BHR REVIEW: WE HAVE REVIEWED OUR PRODUCTION AND INSPECTION RECORDS AND HAVE ESTABLISHED THAT ALL PRODUCTION AND QUALITY PROCESSES WERE CARRIED OUT NORMALLY. NEITHER QA NOR NCMR'S. SYRINGES WERE PACKED IN MACHINE Nº2022 (APRIL 23 - 24TH, 2017). SYRINGES WERE ASSEMBLED IN MACHINE, Nº4251, Nº4252, Nº4204, AND Nº4208, IN LOT #7100072 (APRIL 18 - 24TH, 2017) AND IN LOT #7112280 (APRIL 24TH - MAY 1ST, 2017). RESEARCH HAS FOUND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE. WE HAVE ALSO REVIEWED THE BARREL LOTS #7100036, #7089484, AND #7114312 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. WE HAVE ALSO REVIEWED THE PLUNGER LOTS #7100040, #7089486, #7083123 AND #77114316 AND NO PROBLEMS, DEFECTS OR QN RELATED TO THE REPORTED ISSUE WERE FOUND. ROOT CAUSE ANALYSIS: THIS LUBRICANT IS INCLUDED IN THE PLASTIC FORMULATION AND IT IS INHERENT TO THE DESIGN AND FUNCTION OF 2 PIECE SYRINGES. DURING THE MANUFACTURING PROCESS, THE LUBRICANT FORMS A MICROSCOPIC LAYER IN THE INTERNAL/EXTERNAL WALLS OF THE BARREL. WHEN THE PLUNGER IS MOVED BACKWARDS DURING THE FILLING OF THE SYRINGE, MOST OF THIS MICROSCOPE LAYER OF LUBRICANT IS DRAGGED BEHIND THE PLUNGER AND A SMALL QUANTITY STILL REMAINS INSIDE TO ALLOW A GOOD SLIDING PERFORMANCE DURING THE INJECTION OPERATION. THE PRESENCE PARTICLES OF LUBRICANT IN THE FLUID PATH OF DISCARDIT II SYRINGES HAS BEEN EVALUATED BY BD. THE POLYPROPYLENE USED TO PRODUCE THE SYRINGE BARRELS (WITH THE SLIP AGENT INCLUDED IN THE FORMULATION) HAS PASSED ALL THE BIOCOMPATIBILITY TESTS REQUIRED PRIOR TO MARKETING THE PRODUCT AND MEET THE ESTABLISHED CRITERIA FOR PRECLINICAL TOXICOLOGICAL SAFETY EVALUATIONS. CONFIRMATION:THE RETURNED SAMPLES PRESENTED WHITE PARTICLES IN THE SYRINGE AND CONFIRMED THE REPORTED ISSUE. BASED ON AN EVALUATION OF SEVERITY AND OCCURRENCE IT WAS DETERMINATE THAT NO CORRECTIVE ACTION IS REQUIRED AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(6). PMA / 510(K) #: THERE IS NO 510(K) FOR THIS DEVICE AS IT IS MANUFACTURED OUTSIDE THE US AND NOT SOLD IN THE US A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED DURING USE OF THE 20 ML DISCARDIT II SYRINGE W/O NEEDLE THERE WAS WHITE PLASTIC FOREIGN MATTER ON THE SYRINGE AND PLUNGER. THERE WAS NO REPORT OF INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
754987 20ML DISCARDIT II SYRINGE W/O NEEDLE SYRINGE WITHOUT NEEDLE FMF BECTON DICKINSON, S.A. 1704190

Patients

Seq Age Sex Outcome Treatment
1 Other