FDA Adverse Event Injury Summary report: N

UNIV RNGLC SHL 48MM/L22

MDR report key: 6977114 · Received October 25, 2017

Report

Report Number
0001822565-2017-07465
Event Type
Injury
Date Received
October 25, 2017
Date of Event
July 1, 2016
Report Date
October 25, 2017
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JWH
PMA / PMN Number
PK113369
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: E-POLY 32MM +3 HIWALL LNR SZ22 P/N EP-108322 L/N 821500. CER BIOLOXD OPTION HD 32MM P/N 650-1056 L/N 907600. CER OPTION TYPE 1 TPR SLEVE +3 P/N 650-1067 L/N 816140. REPORTED EVENT WAS CONFIRMED BY REVIEW OF PHOTOS PROVIDED. DHR WAS REVIEWED AND NO DISCREPANCIES WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED. PICTURES WERE REVIEWED WITH HEALTH CARE PROFESSIONAL AND DEVELOPMENT ENGINEER AND CONFIRMED THAT CUP POSITION WAS RETROVERTED. DEFORMATION WAS FOUND NEAR THE FRACTURED PART INDICATING THE HIGH WALL LINER WAS SUBJECTED TO OVER LOADING. IT IS POSSIBLE THAT THE IMPLANT POSITION LED TO IMPINGEMENT AND CAUSING EXCESSIVE WEAR AND LOADING ON THE HIGH WALL LINER RESULTING IN THE FRACTURE AND DISLOCATION OF THE HEAD. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT THIS TYPE OF EVENT CAN OCCUR AND RISKS ARE ADDRESSED IN THE ASSOCIATED RISK DOCUMENTATION. ROOT CAUSE COULD NOT BE DETERMINED WITH INFORMATION AVAILABLE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Description of Event or Problem · 1

PATIENT UNDERWENT A LEFT HIP REVISION PROCEDURE APPROXIMATELY TEN MONTHS POST-IMPLANTATION DUE TO DISLOCATION. DURING THE PROCEDURE, THE ACETABULAR LINER WAS NOTED TO BE FRACTURED. THE FEMORAL HEAD, ACETABULAR LINER AND CUP WERE REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
758223 UNIV RNGLC SHL 48MM/L22 HIP PROSTHESIS JWH ZIMMER BIOMET, INC. N/A 606460

Patients

Seq Age Sex Outcome Treatment
1 76 YR Hospitalization| R