AXIOM ARTIS DTA
Report
- Report Number
- 3004977335-2017-98639
- Event Type
- Malfunction
- Date Received
- October 25, 2017
- Date of Event
- September 29, 2017
- Report Date
- September 29, 2017
- Manufacturer
- SIEMENS HEALTHCARE GMBH
- Product Code
- OWB
- PMA / PMN Number
- K052202
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
EXEMPTION NUMBER E2017017. (B)(4). SIEMENS IS CONDUCTING A THOROUGH INVESTIGATION OF THE REPORTED EVENTS. AS THIS EVENT IS UNDER INVESTIGATION, A ROOT CAUSE HAS NOT YET BEEN DETERMINED. A SUPPLEMENT REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.
IT WAS REPORTED TO SIEMENS THAT A MALFUNCTION OCCURRED WHILE OPERATING THE AXIOM ARTIS DTA SYSTEM. DURING AN INTERVENTIONAL PROCEDURE, THE CUSTOMER REPORTED THAT FLUORO WAS NOT AVAILABLE AND A MICROFOCUS DEFECT HAD OCCURRED. AT THS TIME, THERE IS NO REPORT OF IMPACT TO THE STATE OF HEALTH OF THE PATIENT INVOLVED.
SIEMENS HAS COMPLETED AN INVESTIGATION OF THE REPORTED EVENT. ACCORDING TO THE LOG FILES, THE MICRO FOCUS OF THE AFFECTED X-RAY TUBE SHOWED ERRORS FOR SOME TIME. AT EACH SYSTEM RESTART, THE GENERATOR DETECTED THE ISSUE, DISABLED THE MICRO FOCUS AND A CORRESPONDING MESSAGE WAS DISPLAYED TO THE USER. IN SUCH CASES, X-RAY IMAGING IS STILL POSSIBLE WITH THE SMALL OR LARGE FOCUS. IF AN ORGAN PROGRAM IS SELECTED USING THE MICRO FOCUS, THE SYSTEM AUTOMATICALLY SWITCHES TO THE SMALL FOCUS. THIS RESULTS IN A MODERATE LIMITATION OF THE IMAGE QUALITY WITH REGARD TO SPATIAL RESOLUTION. THE ROOT CAUSE OF THE MICRO FOCUS ISSUE COULD NOT BE DETERMINED. THE X-RAY TUBE WAS INSTALLED IN 2011. THUS, A POSSIBLE ROOT CAUSE FOR THE MICRO FOCUS ISSUE IS A DEFECT DUE TO WEAR OF THE COMPONENT. THE USER REBOOTED THE SYSTEM RESETTING AND NEWLY INITIALIZING ALL CAN COMPONENTS. AFTER THIS, THE SYSTEM RECOVERED AND THE FULL X-RAY IMAGING FUNCTIONALITY WAS AVAILABLE AGAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 755734 | AXIOM ARTIS DTA | INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM | OWB | SIEMENS HEALTHCARE GMBH | 7008605 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |