FDA Adverse Event Malfunction Summary report: N

DADE ACTIN FSL ACTIVATED PTT REAGENT

MDR report key: 6973814 · Received October 24, 2017

Report

Report Number
9610806-2017-00117
Event Type
Malfunction
Date Received
October 24, 2017
Date of Event
September 28, 2017
Report Date
October 24, 2017
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Product Code
GGW
UDI-DI
00842768003882
PMA / PMN Number
K863594
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS HEALTHCARE DIAGNOSTICS INC. (SIEMENS) HEADQUARTERS SUPPORT CENTER (HSC) SPECIALIST REVIEWED THE DATA PRINTOUTS AND INTERNAL QUALITY CONTROLS TO DETERMINE THE CAUSE OF FLAGGED, NON-NUMERICAL RESULTS ON THE SYSMEX CS-2500 SYSTEM AND THE SYSMEX CS-5100 SYSTEM. THE HSC SPECIALIST DETERMINED THAT THE INTERNAL QUALITY CONTROLS RECOVERED WITHIN RANGE ON BOTH SYSTEMS AND THE ISSUE IS ISOLATED TO TWO SAMPLES FROM THE AFFECTED PATIENT. THE HSC SPECIALIST DETERMINED THAT THE CUSTOMER FOLLOWED SIEMENS'S PROCEDURE AND GOOD LABORATORY PRACTICES WHEN EVALUATING THE ABNORMAL AND FLAGGED RESULTS. THE RESULTS OBTAINED USING SAMPLE (B)(6) WERE NOT REPORTABLE AND THE CUSTOMER DID NOT REPORT THESE RESULTS. THE CUSTOMER APPROPRIATELY REQUESTED FOR A REDRAWN PATIENT SAMPLE. THE KINETICS OF THE MEASURING CURVE OBTAINED AT 21:38 USING SAMPLE (B)(6) INDICATED A CLEAR BASELINE, CLOTTING AND ENDPOINT PHASE OF THE REACTION. THE KINETICS OF THIS MEASUREMENT CURVE SUPPORTS THAT THE RESULT FOR THE AFFECTED SAMPLE WAS GREATER THAN THE REPORTABLE LIMIT OF 139 SECONDS. THIS RESULT ALIGNS WITH THE PATIENT'S UNFRACTIONATED (UF) HEPARIN THERAPY AND THE RESULT OBTAINED AT THE ALTERNATE FACILITY. THE PATIENT WAS ADMINISTERED UF HEPARIN WITH AN IV AND UF HEPARIN IS AN ANTICOAGULANT THAT CAUSES ELEVATION OF THE ACTIVATED PARTIAL THROMBOPLASTIN TIME (APTT) RESULTS. IF THE PATIENT IS ADMINISTERED UF HEPARIN THERAPY VIA AN IV, THE PATIENT'S BLOOD MUST BE COLLECTED SUCH THAT THE UF HEPARIN INFUSION DOES NOT CONTAMINATE THE SAMPLE. SIEMENS IS UNABLE TO DETERMINE IF THE SAMPLES WERE COLLECTED CORRECTLY. HOWEVER, THE CUSTOMER INDICATED THAT THE AFFECTED SAMPLES WERE PERIPHERALLY COLLECTED AND THE SAMPLES' INTEGRITY WERE ACCEPTABLE. THE SAME DADE ACTIN FSL ACTIVATED PTT REAGENT LOTS WERE USED ON BOTH SYSTEMS, EXCLUDING A GENERAL ISSUE WITH THE SYSTEM AND REAGENT USED. THE CAUSE OF THE FLAGGED, NON-NUMERICAL APTT RESULTS IS UNKNOWN. THE SYSTEMS AND REAGENT ARE PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THESE SYSTEMS OR REAGENT IS REQUIRED.

Description of Event or Problem · 1

A FLAGGED, NON-NUMERICAL ACTIVATED PARTIAL THROMBOPLASTIN TIME (APTT) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON THE SYSMEX CS-5100 SYSTEM. THE PATIENT SAMPLE WAS RERUN WITH EXTENDED MEASUREMENT TIME ON THE SAME SYSTEM AND A NON-NUMERICAL, FLAGGED RESULT WAS OBTAINED. THE PATIENT'S BLOOD WAS REDRAWN AND RUN, IN DUPLICATE, ON THE SAME SYSTEM AND NON-NUMERICAL FLAGGED RESULTS WERE OBTAINED. THE REDRAWN PATIENT SAMPLE WAS RERUN ON AN ALTERNATE SYSMEX CS-2500 SYSTEM, RESULTING IN A NON-NUMERICAL FLAGGED RESULT. THE CUSTOMER SENT OUT THE PATIENT SAMPLE TO AN ALTERNATE LAB, WHERE THE SAMPLE WAS RERUN ON A BCS XP SYSTEM, RESULTING IN AN ELEVATED APTT RESULT. THE CUSTOMER REPORTED THIS RESULT AS >200 SECONDS TO THE PHYSICIAN AND THE PHYSICIAN MODIFIED THE PATIENT'S HEPARIN THERAPY BASED ON THIS RESULT. THE CUSTOMER ALLEGED THAT THE DELAY IN REPORTING OF RESULTS FROM THE SYSMEX CS-5100 SYSTEM AND SYSMEX CS-2500 SYSTEM CAUSED THE PATIENT THAT HAD UNDERGONE A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE TO BLEED FROM HIS WOUND. THE BLEEDING WAS STOPPED WITH APPLIED PRESSURE TO THE RIGHT GROIN FOR AN HOUR. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DELAY IN REPORTING OF APTT RESULTS ON THE PATIENT SAMPLES.