FDA Adverse Event Injury Summary report: N

530G INSULIN PUMP MMT-551LNAS

MDR report key: 6971873 · Received October 24, 2017

Report

Report Number
2032227-2017-59969
Event Type
Injury
Date Received
October 24, 2017
Date of Event
October 9, 2017
Report Date
February 27, 2018
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
UDI-DI
00643169507739
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. INSULIN PUMP PASSED ALL FUNCTIONAL TESTING, INCLUDING IDLE CURRENT TEST, RUN CURRENT TEST, SELF-TEST, OFF NO POWER TEST, UNEXPECTED RESTART ERROR TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME/COMPROMISED FORCE SENSOR SYSTEM TEST, EXCESSIVE NO DELIVERY TEST, DISPLACEMENT TEST, DISPLACEMENT ACCURACY TEST AT 0.08750 INCHES AND REWIND TEST. NO OVER DELIVERY ANOMALY NOTED. INSULIN PUMP HAD MINOR SCRATCHED DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE LEVELS AND OVER DELIVERY. THE CUSTOMER HAD SYMPTOMS SUCH AS ANXIETY, SWEATING, AND CONFUSION. THE CUSTOMER WAS TREATED FOR THE LOW BLOOD GLUCOSE WITH APPLE JUICE AND PBJ SANDWICH. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 43 MG/DL AT THE TIME OF THE INCIDENT AND CURRENT BLOOD GLUCOSE LEVELS WAS 196 MG/DL. CUSTOMER STATED THAT THE INSULIN PUMP WAS OVER DELIVERING INSULIN. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDS TO BE REPLACED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
750884 530G INSULIN PUMP MMT-551LNAS ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-551LNAS A4551LNASJ 00643169507739

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other