EKOSONIC ENDOVASCULAR CATHETER

MDR report key: 6969780 · Received October 23, 2017

Report

Report Number
3001627457-2017-00029
Event Type
Malfunction
Date Received
October 23, 2017
Date of Event
September 17, 2017
Report Date
September 26, 2017
Manufacturer
EKOS CORPORATION
Product Code
KRA
UDI-DI
00858593006318
PMA / PMN Number
K140151
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE EVENT LOG CONFIRMED THE REPORTED ALL MSD GROUPS DISABLED ALARM 36 MINUTES INTO THE TREATMENT. THE INVESTIGATION OF THE RETURNED DEVICE ON 26 SEPTEMBER 2017 CONFIRMED THE MSD SHAFT WAS FRACTURED. THE MSD SHAFT WAS DAMAGED WITH THE RF COMMON WIRE BROKEN, WHICH CAUSED THE MSD FAILURE. THE OPTICAL MICROSCOPY EVALUATION NOTED A PINCH MARK ON THE MSD SHAFT AND DAMAGE 3CM FROM THE END OF THE LUER BARB WITH EXPOSING WIRES INSIDE. IT IS LIKELY THAT THE COMPLAINANT USED A PLACEMENT AID OR TOOL DURING MSD PLACEMENT WHICH CONTRIBUTED TO THE DAMAGE. REVIEW OF THE DEVICE HISTORY RECORD INDICATED THAT THE DEVICE WAS MANUFACTURED ACCORDING TO SPECIFICATION AND MET ALL ACCEPTANCE CRITERIA PRIOR TO RELEASE. THE INSTRUCTIONS FOR USE WARN THE USER "DO NOT DEFORM OR KINK THE MSD DURING DELIVERY INTO THE IDDC. IF THE MSD IS KINKED AT ANY TIME, DO NOT ATTEMPT TO USE THE MSD, AS KINKING MAY LEAD TO DEGRADED PERFORMANCE OR FRACTURE DURING USE." NO SERIOUS INJURY WAS REPORTED. THE PATIENT WAS REPORTED AS DOING WELL.

Description of Event or Problem · 1

ON 17 SEPTEMBER 2017, IT WAS REPORTED THAT THE DEVICE WAS PLACED SUCCESSFULLY FOR A LEFT LEG ARTERIAL CASE WITH CONTRALATERAL ACCESS BUT FAILED AFTER 44 MINUTES WITH THE ALARM CODE - ALL MSD GROUPS DISABLED. ALL FLUIDS WERE LEFT RUNNING AND THE PATIENT WAS REPORTED AS DOING WELL. NO FURTHER INTERVENTION WAS REPORTED. THE CATHETER WAS RETURNED FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
749256 EKOSONIC ENDOVASCULAR CATHETER CATHETER KRA EKOS CORPORATION 500-56150 00858593006318

Patients

Seq Age Sex Outcome Treatment
1