FDA Adverse Event Malfunction Summary report: N

BARDEX® I.C. FOLEY CATHETER

MDR report key: 6969629 · Received October 23, 2017

Report

Report Number
1018233-2017-05402
Event Type
Malfunction
Date Received
October 23, 2017
Date of Event
September 25, 2017
Report Date
November 14, 2017
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
MJC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE LOT NUMBER IS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE INSTRUCTIONS FOR USE STATE THE FOLLOWING: "INDICATIONS FOR USE: THE BARDEX® I.C. LATEX FOLEY CATHETER IS INDICATED FOR USE IN THE DRAINAGE AND/OR COLLECTION AND/OR MEASUREMENT OF URINE. GENERALLY, DRAINAGE IS ACCOMPLISHED BY INSERTING THE CATHETER THROUGH THE URETHRA AND INTO THE BLADDER. HOWEVER, DRAINAGE IS SOMETIMES ACCOMPLISHED BY SUPRAPUBIC OR OTHER PLACEMENT OF THE CATHETER, SUCH AS NEPHROSTOMY TRACT. DIRECTIONS FOR USE: VISUALLY INSPECT THE PRODUCT FOR ANY IMPERFECTIONS OR SURFACE DETERIORATION PRIOR TO USE. IF PACKAGE IS OPENED OR IF ANY IMPERFECTION OR SURFACE DETERIORATION IS OBSERVED, DO NOT USE. USING PROPER ASEPTIC METHODS, REMOVE CATHETER FROM PACKAGE. PREPARE PATIENT PER HOSPITAL/NURSING RECOMMENDED PROCEDURE. PROCEED WITH CATHETERIZATION USING STANDARD TECHNIQUES. INFLATE THE BALLOON WITH STERILE WATER TO THE VOLUME STATED ON THE INFLATION LUMEN OF THE CATHETER BY USING A WATER-FILLED LUER TIP SYRINGE. CONNECT CATHETER TO COLLECTION CONTAINER. TO DEFLATE THE BALLOON, GENTLY REINSERT THE LUER TIP SYRINGE INTO THE VALVE AND ASPIRATE. CAUTION: TO DEFLATE CATHETER BALLOON: GENTLY INSERT A LUER SLIP TIP SYRINGE IN THE CATHETER VALVE. NEVER USE MORE FORCE THAN IS REQUIRED TO MAKE THE SYRINGE ¿STICK¿ IN THE VALVE. ALLOW THE PRESSURE IN THE BALLOON TO FORCE THE PLUNGER BACK AND FILL THE SYRINGE WITH WATER. IF YOU NOTICE SLOW OR NO DEFLATION, RE-SEAT THE SYRINGE GENTLY. USE ONLY GENTLE ASPIRATION TO ENCOURAGE DEFLATION IF NEEDED. VIGOROUS ASPIRATION MAY COLLAPSE THE INFLATION LUMEN, PREVENTING BALLOON DEFLATION. IF PERMITTED BY HOSPITAL PROTOCOL, THE VALVE ARM MAY BE SEVERED. IF THIS FAILS, CONTACT ADEQUATELY TRAINED PROFESSIONAL FOR ASSISTANCE, AS DIRECTED BY HOSPITAL PROTOCOL. SHOULD BALLOON RUPTURE OCCUR, CARE SHOULD BE TAKEN TO ASSURE THAT ALL BALLOON FRAGMENTS HAVE BEEN REMOVED FROM THE PATIENT. THIS IS A SINGLE USE DEVICE. DO NOT RESTERILIZE ANY PORTION OF THIS DEVICE. REUSE AND/OR REPACKAGING MAY CREATE A RISK OF PATIENT OR USER INFECTION, COMPROMISE THE STRUCTURAL INTEGRITY AND/OR ESSENTIAL MATERIAL AND DESIGN CHARACTERISTICS OF THE DEVICE, WHICH MAY LEAD TO DEVICE FAILURE, AND /OR LEAD TO INJURY, ILLNESS OR DEATH OF THE PATIENT. VALVE TYPE: USE LUER SLIP TIP SYRINGE. DO NOT USE NEEDLE. DO NOT EXCEED RECOMMENDED CAPACITIES AS STATED ON THE INFLATION LUMEN OF THE CATHETER. CAUTION: FEDERAL (U.S.A.) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. STORAGE: STORE CATHETERS AT ROOM TEMPERATURE AWAY FROM DIRECT EXPOSURE TO LIGHT, PREFERABLY IN THE ORIGINAL BOX. CAUTION: DO NOT ASPIRATE URINE THROUGH DRAINAGE FUNNEL. WARNING: ON CATHETER, DO NOT USE OINTMENTS OR LUBRICANTS HAVING A PETROLATUM BASE. THEY WILL DAMAGE LATEX AND MAY CAUSE BALLOON TO BURST. AFTER USE, THIS PRODUCT MAY BE A POTENTIAL BIOHAZARD. HANDLE AND DISPOSE OF IN ACCORDANCE WITH ACCEPTED MEDICAL PRACTICE AND APPLICABLE LOCAL, STATE AND FEDERAL LAWS AND REGULATIONS. CAUTION: THIS PRODUCT CONTAINS NATURAL RUBBER LATEX WHICH MAY CAUSE ALLERGIC REACTIONS. STERILE UNLESS PACKAGE IS OPENED OR DAMAGED. DO NOT USE IF PACKAGE IS DAMAGED. RECOMMENDED INDWELLING TIME NOT TO EXCEED 28 DAYS." (B)(4). THE DEVICE WAS NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CATHETER'S DIAMETER LOOKED THINNER, ABOUT 14 FR, (TW 1053211 ) THAN PRODUCT CATALOG NO.120716, A 16FR. THE USER OPENED ANOTHER CATHETER THAT ALSO APPEARED THINNER (TW #1063736). THE THIRD ONE WAS OPENED, AND IT LOOKED LIKE THE CORRECT DIAMETER (16FR), SO THE USER WAS ABLE TO USE THE THIRD ONE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CATHETER'S DIAMETER LOOKED THINNER, ABOUT 14 FR, THAN PRODUCT CATALOG NO.120716, A 16FR. THE USER OPENED ANOTHER CATHETER THAT ALSO APPEARED THINNER. THE THIRD ONE WAS OPENED, AND IT LOOKED LIKE THE CORRECT DIAMETER (16FR), SO THE USER WAS ABLE TO USE THE THIRD ONE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
747774 BARDEX® I.C. FOLEY CATHETER FOLEY CATHETER (LATEX) MJC C.R. BARD, INC. (COVINGTON) -1018233 UNK

Patients

Seq Age Sex Outcome Treatment
1