FDA Adverse Event Other Summary report: N

ANGENIEUX

MDR report key: 69682 · Received February 12, 1997

Report

Report Number
2027060-1997-09003
Event Type
Other
Date Received
February 12, 1997
Date of Event
December 19, 1996
Report Date
February 12, 1997
Manufacturer
ALM HOSP EQUIPMENT
Product Code
FQP
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
MD, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE, THE FINISH PAINT ON A LIGHT ARM FLAKED AND POSSIBLY FELL INTO THE OPEN WOUND OF A PT. THE HOSP INFORMED MFR TECH THAT THE PT SUFFERED NO INFECTION FROM THE PAINT. UPON INVESTIGATION, THE TECH FOUND A LOCATION THAT HAD CHIPPED PAINT. THE CHIPPED FORK WAS REPLACED BY A NEW ONE. AFTER INSPECTION BY MFR'S QUALITY CONTROL, IT WAS DETERMINED THAT THE CHIPPING WAS CAUSED BY THE IMPACT OF TWO LIGHTHEADS HIT TOGETHER BY HOSP PERSONNEL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANGENIEUX SURGICAL LIGHTING FQP ALM HOSP EQUIPMENT 310013 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other