Additional Manufacturer Narrative · 1

THE DEVICE INVOLVED IN THE REPORTED INCIDENT IS NOT AVAILABLE FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.

Additional Manufacturer Narrative · 1

ON 10/24/17 INTEGRA INVESTIGATION COMPLETED. METHOD: FAILURE ANALYSIS, DEVICE HISTORY EVALUATION. COMPLAINT IS UNCONFIRMED; THIS IS DUE TO THE PRODUCT NOT BEING RETURNED FOR REVIEW. DEVICE HISTORY EVALUATION - DHR REVIEW NONCONFORMING PRODUCT REPORT/ NONCONFORMING MATERIAL REPORT HISTORY: NONE VARIANCE AUTHORIZATION/DEVIATION HISTORY: NONE. ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY: THERE IS NO APPLICABLE ENGINEERING CHANGE ORDER/MANUFACTURING CHANGE ORDER HISTORY: CORRECTIVE ACTION PREVENTIVE ACTION HISTORY: NONE. HEALTH HAZARD EVALUATION HISTORY: NONE. ROOT CAUSE CANNOT BE DETERMINED DUE TO THE LACK OF INFORMATION RECEIVED TO PERFORM A COMPLETE INVESTIGATION. PRODUCT HAS NOT BEEN RETURNED FOR EVALUATION.

Description of Event or Problem · 1

FDA SUS VOLUNTARY EVENT REPORT MW5072008 REPORTS THAT THE METAL TIP OF THE SURGICAL JARIT CARB EDGE SCISSOR #101267 BROKE OFF DURING AN ABDOMINAL SURGICAL PROCEDURE X-RAY WERE NEGATIVE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
747249 CARB-EDGE METZ SCS 9 CVD CARDIOVASCULAR SCISSORS DWS INTEGRA YORK, PA INC. UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1