FDA Adverse Event Injury Summary report: N

GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET

MDR report key: 6964403 · Received October 20, 2017

Report

Report Number
3002808486-2017-02082
Event Type
Injury
Date Received
October 20, 2017
Report Date
March 1, 2019
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
PMA / PMN Number
K090140
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI), 400 DANIELS WAY, BLOOMINGTON, IN 47404, REGISTRATION NO.: 3005580113. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW UP REPORT WILL BE SUBMITTED SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURER REF# (B)(4). D4) CATALOG#: UNKNOWN BUT REFERRED TO AS A GUNTHER TULIP FILTER. G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI) 400 DANIELS WAY BLOOMINGTON, IN 47404 REGISTRATION NO.: 3005580113. G5) SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K090140, K112119 OR K12105. (B)(4). IT HAS NOT BEEN POSSIBLE TO INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVING WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56.

Description of Event or Problem · 0

PER CT, (B)(6) 2019: IVC FILTER NOTED. THERE IS PERFORATION OF THE IVC FILTER STRUT COMPONENTS BEYOND THE WALL OF THE IVC. THERE IS PERFORATION INTO THE MESENTERY SECONDARY TO THE PERFORATION OF THE IVC FILTER STRUT COMPONENTS BEYOND THE WALL OF THE IVC.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK GUNTHER TULIP FILTER ON (B)(6) 2012". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016032. (B)(4). NAME AND ADDRESS FOR IMPORTER SITE: (B)(4). THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016032. (B)(4). CORRECTED DATA BASED ON NEW INFORMATION RECEIVED: ADVERSE EVENT TO ADVERSE EVENT AND PRODUCT PROBLEM. IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'VENA CAVA PERFORATION, ORGAN PERFORATION (MESENTERY), PULMONARY EMBOLISM, ANXIETY'. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. VENA CAVA WALL PERFORATION IS A KNOWN POTENTIAL COMPLICATION OF VENA CAVA FILTERS. BOTH SYMPTOMATIC AND ASYMPTOMATIC EVENTS HAVE BEEN REPORTED. AMONG OTHER CAUSES, VENA CAVA WALL PERFORATION MAY INADVERTENTLY BE INITIATED BY IMPROPER DEPLOYMENT, EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL PROCEDURE IN THE VICINITY OF A FILTER) AND (OR) PROCEDURES THAT INVOLVE OTHER DEVICES BEING PASSED THROUGH AN IN SITU FILTER. THERE IS A CURRENT DEBATE IN THE PUBLISHED SCIENTIFIC LITERATURE ON A DIFFERENTIATION BETWEEN IVC WALL PERFORATION WITH AND WITHOUT CLINICAL SEQUELAE. E.G. FILTER LEGS MAY BE OUTSIDE THE CONTRAST LUMEN ON IMAGING WITHOUT ACTUALLY PERFORATING THE IVC WALL (KNOWN AS TENTING) AND WITH NO CLINICAL SEQUELAE. IN CONTRAST, PERFORATION OF ADJACENT ORGANS IS REPORTED WITH CLINICAL SEQUELAE. PE IS A KNOWN RISK IN RELATION TO FILTER IMPLANT REPORTED IN THE PUBLISHED SCIENTIFIC LITERATURE. ALSO, IT IS REPORTED THAT THE PULMONARY EMBOLISM IN SOME CASES MAY ORIGINATE FROM UPPER EXTREMITIES INSTEAD OF LOWER EXTREMITY VEINS. WITH ALL FILTERS, THERE IS SOME RISK OF FURTHER PULMONARY EMBOLISM. UNKNOWN IF THE REPORTED ANXIETY IS DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY CORRESPONDING FAILURE MODE(S) AT THIS TIME. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION RECEIVED ON 10/11/2017 AS FOLLOWS: PATIENT RECEIVED AN IMPLANT ON (B)(6) 2012 VIA THE DUE TO DEEP VEIN THROMBOSIS, PULMONARY EMBOLISM, GASTROINTESTINAL BLEED. PATIENT IS ALLEGING VENA CAVA PERFORATION, ORGAN PERFORATION (MESENTERY), PULMONARY EMBOLISM, ANXIETY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
744547 GUNTHER TULIP NAVALIGN FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE E2892593

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening