HEARTSINE SAMARITAN 500P AND PAD PAK
Report
- Report Number
- 3004123209-2017-01206
- Event Type
- Malfunction
- Date Received
- October 20, 2017
- Date of Event
- September 27, 2017
- Report Date
- January 25, 2018
- Manufacturer
- HEARTSINE TECHNOLOGIES LTD
- Product Code
- MKJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- PATIENT
Narratives
EXEMPTION NUMBER E2015058. (B)(4). ON RECEIPT OF THE DEVICE THE HISTORY AND MEMORY LOGS WERE DOWNLOADED. THE HISTORY LOG FOR THIS DEVICE SHOWED THAT THE PAD-PAK WAS FIRST INSTALLED ON THE (B)(6) 2014 AND THAT IT HAD PERFORMED TO ALL WEEKLY AUTO SELF-TESTS UP TO AND INCLUDING THE LAST LOG ENTRY, PRIOR TO RECEIPT AT HEARTSINE, ON THE (B)(6) 2017. THE RETURNED SAM 500P PASSED "OUT QAT" FROM HEARTSINE TECHNOLOGIES ON THE (B)(6) 2013. THE DEVICE WAS TESTED ON THE CALIBRATED IMPULSE DEFIBRILLATOR ANALYSER USING THE RETURNED PAD-PAK AND THE DEVICE DELIVERED A TEST SHOCK WITHOUT FAULT. PREVIOUS EXPERIENCE HAS SHOWN THAT A LEAKAGE CURRENT ACROSS ONE OF THE TRANSISTORS IN THE CIRCUIT MAY RESULT IN UNCHARACTERISTIC LIGHTING OF THE STATUS LEDS AND COULD POTENTIALLY CAUSE BOTH LEDS TO FLASH SIMULTANEOUSLY. THE DEVICE WAS DISASSEMBLED TO INVESTIGATE AND THE LED DRIVE WAS SCRUTINIZED WITH NO MEASURABLE FAULT FOUND. AN INTERMITTENT COMPONENT FAILURE MAY HAVE RESULTED IN THE REPORTED FAULT AND AN EXCESS CURRENT DRAIN. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE
EXEMPTION NUMBER E2015058. HEARTSINE TECHNOLOGIES LTD (MANUFACTURER) IS SUBMITTING THE REPORT ON BEHALF OF HEARTSINE TECHNOLOGIES LLC (IMPORTER).
THERE WAS NO PATIENT INVOLVED IN THIS EVENT. RED AND GREEN STATUS INDICATORS DISPLAYED SIMULTANEOUSLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 744508 | HEARTSINE SAMARITAN 500P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES LTD | SAM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |