FDA Adverse Event Malfunction Summary report: N

PROTEUSPLUS

MDR report key: 6961204 · Received October 19, 2017

Report

Report Number
3000256071-2017-00003
Event Type
Malfunction
Date Received
October 19, 2017
Date of Event
September 27, 2017
Report Date
September 27, 2017
Manufacturer
IBA (ION BEAM APPLICATIONS)
Product Code
LHN
PMA / PMN Number
K082416
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY IS PROVIDED.

Description of Event or Problem · 1

DESCRIPTION OF THE EVENT: IBA HAS BEEN INFORMED THAT, ON (B)(6) 2017, DURING A PATIENT TREATMENT, THE RANGE SHIFTER WAS NOT CORRECTLY ALIGNED ON THE RAILS. A RANGE SHIFTER IS USED IN PENCIL BEAM SCANNING TO SHIFT THE DOSE DISTRIBUTION IN THE TARGETED AREA IN THE PATIENT. TREATMENT FIELDS WERE DELIVERED WITH A SLIGHTLY ANGLED RANGE SHIFTER. AFTER THE DELIVERY OF THE TREATMENT, WHILE THE GANTRY WAS AT 340°, THE PATIENT HIT THE RANGE SHIFTER WHILE GETTING SITTED. THE RANGE SHIFTER (23 POUNDS) THEN FELL OUT AND GRAZED THE PATIENT FACE. IMPACT OF THE EVENT ON THE PATIENT: THE PATIENT WAS SENT TO THE EMERGENCY DEPARTMENT OF THE HOSPITAL FOR EVALUATION. THE PATIENT WAS RELEASED WITH NO INJURIES. THE HOSPITAL EVALUATED THE IMPACT OF THE SLIGHTLY ANGLED RANGE SHIFTER ON BEAM PROPERTIES IN TREATMENT FIELD TO LESS THAN 1 MM IN WATER-EQUIVALENT-THICKNESS CHANGE FOR THE TREATMENT FIELD, THEREFORE WELL WITHIN THEIR MARGIN USED IN TREATMENT PLANNING. IBA IS PROACTIVELY REPORTING THIS EVENT DUE TO THE HEAVY WEIGHT OF THE RANGE SHIFTER AND SIGNIFICANT IMPACT IT COULD HAVE IF THERE WAS A NEW OCCURRENCE. EVALUATION SUMMARY: INVESTIGATION DEMONSTRATED THAT THE RANGE SHIFTER CAN BE INSERTED WITH ONLY THE UPPER RAIL IN THE GUIDE, THE LOWER RAIL RESTING JUST OUTSIDE THE GUIDE (SEE ATTACHMENT). WITH THIS CONFIGURATION, THE LOCKING PIN WILL LOCK THE RANGE SHIFTER INTO PLACE AND THE INSERTION PINS WILL ALSO BE PUSHED IN (SEE ATTACHMENTS). HOWEVER, WITH MINIMAL OUTWARD PRESSURE, THE RANGE SHIFTER CAN BE PULLED OUT OF THE HOLDER AND FALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
741771 PROTEUSPLUS PROTON THERAPY SYSTEM LHN IBA (ION BEAM APPLICATIONS) PROTEUS 235

Patients

Seq Age Sex Outcome Treatment
1 Other