FDA Adverse Event
Injury
Summary report: N
BERCHTOLD OPERATING TABLE
MDR report key: 696072
·
Received March 23, 2006
Report
- Report Number
- MW1038340
- Event Type
- Injury
- Date Received
- March 23, 2006
- Date of Event
- August 1, 2005
- Report Date
- March 23, 2006
- Manufacturer
- BERCHTOLD CORP
- Product Code
- FQO
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
OPERATING TABLE DEFECT. DURING A PRONE CRAINOTOMY THE BED WAS LOWERED FOR BETTER VISUALIZATION FOR THE SURGEON USING A MICROSCOPE. UNKNOWN AT THE TIME THE TELESCOPIC BASE OF THE BED WAS NOT INTACT. AS SUCH, THE ANESTHESIA CIRCUIT SLIPPED BETWEEN THE TELESCOPIC PORTION OF THE FRAME AND THE FIXED PLATFORM. AS THE BED WAS LOWERED THE ANESTHESIA CIRCUIT WAS SEVERED IN HALF. THE PT WAS QUICKLY OXYGENATED WITH BACKUP EQUIPMENT UNTIL A REPLACEMENT WAS IN-LINE. SHE SUFFERED NO ILL EFFECTS FROM THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BERCHTOLD OPERATING TABLE | OPERON SCANDIA SL+ | FQO | BERCHTOLD CORP | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Disability |