ENDO GIA
Report
- Report Number
- 1219930-2017-08048
- Event Type
- Malfunction
- Date Received
- October 18, 2017
- Date of Event
- January 6, 2017
- Report Date
- October 18, 2017
- Manufacturer
- COVIDIEN LP LLC NORTH HAVEN
- Product Code
- GDW
- PMA / PMN Number
- K111825
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
EVALUATION SUMMARY: POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE DEVICE. VISUAL INSPECTION OF THE CARTRIDGE REVEALED THAT THE LOADING UNIT WAS PARTIALLY FIRED. THE ANVIL OF THE LOADING UNIT WAS SEVERELY BOWED AND THE KNIFE BAR ASSEMBLY WAS BUCKLED. DUE TO THE NOTED DAMAGE NO FUNCTIONAL TESTING WAS PERFORMED ON THE LOADING UNIT. RECORDS FROM EACH MANUFACTURING LOT ARE THOROUGHLY REVIEWED TO ENSURE THAT PRODUCTS ARE RELEASED MEETING ALL QUALITY RELEASE SPECIFICATIONS AT THE TIME OF MANUFACTURE. REPLICATION OF THE BOWED ANVIL, DEFORMED ANVIL CLAMPING MECHANISM, AND BUCKLED KNIFE-BAR ASSEMBLY MAY OCCUR UNDER THE FOLLOWING CONDITIONS: FIRING OVER TISSUE THAT IS BEYOND THE RECOMMENDED THICKNESS RANGE, FIRING WITH AN OBSTACLE INCORPORATED IN THE JAWS. IN EITHER OF THESE CIRCUMSTANCES, IT WILL BECOME INCREASINGLY DIFFICULT TO ACTUATE THE FIRING HANDLE AND THE INSTRUMENT RETURN KNOBS WILL BE DIFFICULT TO RETRACT. IN ADDITION, STAPLES MAY NOT FORM PROPERLY AND TISSUE MAY NOT BE FULLY TRANSECTED. THE ROOT CAUSE OF THE OBSERVED DAMAGE WAS MISUSE OF THE PRODUCT WHICH WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED INCIDENT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
ACCORDING TO THE REPORTER, DURING THE PROCEDURE VATS LOBECTOMY. THE STAPLER CAN'T WORK WELL (WIDTH OF JAW OPENING IS SMALLER THAN USUAL) WHEN SURGEON RESECTION THE LOBE. SO THE SURGEON CHANGED TO A NEW ONE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 737989 | ENDO GIA | STAPLE, IMPLANTABLE | GDW | COVIDIEN LP LLC NORTH HAVEN | 030458 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |