FDA Adverse Event Injury Summary report: N

COOK CELECT FEMORAL VENA CAVA FILTER SET

MDR report key: 6957272 · Received October 18, 2017

Report

Report Number
3002808486-2017-01995
Event Type
Injury
Date Received
October 18, 2017
Report Date
August 11, 2022
Manufacturer
WILLIAM COOK EUROPE
Product Code
DTK
PMA / PMN Number
K073374
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
ATTORNEY
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURER REF# (B)(4). D4) CATALOG#: UNKNOWN BUT REFERRED TO AS A COOK CELECT FILTER. G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI) 400 DANIELS WAY BLOOMINGTON, IN 47404 REGISTRATION NO.: 3005580113. G5) SINCE CATALOG# IS UNKNOWN THE 510(K) COULD BE EITHER K061815, K073374 OR K09014. (B)(4). INVESTIGATION IS STILL IN PROGRESS.

Additional Manufacturer Narrative · 0

BLANK FIELDS ON THIS FORM INDICATE THE INFORMATION IS UNKNOWN, UNAVAILABLE, OR UNCHANGED. CORRECTED DATA: A2 THE FOLLOWING FIELDS WERE UPDATED PER ADDITIONAL INFORMATION RECEIVED: A4, B1, B2, B5, B6, ANNEX E, ANNEX A, ANNEX G, ANNEX B, ANNEX C, ANNEX D, AND H6. INVESTIGATION THE FOLLOWING ALLEGATIONS HAVE BEEN INVESTIGATED: VENA CAVA PERFORATION, TILT, PAIN, DISCOMFORT. INVESTIGATION IS REOPENED DUE TO ADDITIONAL INFORMATION PROVIDED. THE REPORTED ALLEGATIONS HAVE BEEN FURTHER INVESTIGATED BASED ON THE INFORMATION PROVIDED TO DATE. FILTER INTERACTS WITH IVC WALL, E.G. PENETRATION/PERFORATION/EMBEDMENT. THIS MAY BE EITHER SYMPTOMATIC OR ASYMPTOMATIC. POTENTIAL CAUSES MAY INCLUDE IMPROPER DEPLOYMENT; AND (OR) EXCESSIVE FORCE OR MANIPULATIONS NEAR AN IN-SITU FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER). POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: VASCULAR TRAUMA, VENA CAVA PERFORATION, VENA CAVA PENETRATION. FILTER TILT HAS BEEN REPORTED. POTENTIAL CAUSES MAY INCLUDE FILTER PLACEMENT IN IVCS WITH DIAMETERS LARGER THAN THOSE SPECIFIED IN THESE INSTRUCTIONS FOR USE; IMPROPER DEPLOYMENT; MANIPULATIONS NEAR AN IMPLANTED FILTER (E.G., A SURGICAL OR ENDOVASCULAR PROCEDURE IN THE VICINITY OF A FILTER); AND (OR) A FAILED RETRIEVAL ATTEMPT. EXCESSIVE FILTER TILT MAY CONTRIBUTE TO DIFFICULT OR FAILED RETRIEVAL; VENA CAVA WALL PENETRATION/PERFORATION; AND (OR) RESULT IN LOSS OF FILTER EFFICIENCY. POTENTIAL ADVERSE EVENTS THAT MAY OCCUR INCLUDE, BUT ARE NOT LIMITED TO, THE FOLLOWING: UNACCEPTABLE FILTER TILT. UNKNOWN IF THE REPORTED PAIN AND DISCOMFORT IS DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS POINT IN TIME. 20 DEVICES IN LOT. NO RELEVANT NOTES ON WORK ORDER. THE PRODUCT IS MANUFACTURED AND INSPECTED ACCORDING TO SPECIFICATIONS. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS, AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. THIS REPORT INCLUDES INFORMATION KNOWN AT THIS TIME. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED IF ADDITIONAL RELEVANT INFORMATION BECOMES AVAILABLE. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED, THAT A DEATH OR SERIOUS INJURY OCCURRED, OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO SHORT FORM COMPLAINT FILED: IT IS ALLEGED THAT "[PT] RECEIVED A COOK CELECT FILTER ON (B)(6) 2010". PATIENT OUTCOME: IT IS ALLEGED THAT [PT] WAS INJURED WITHOUT FURTHER EXPLANATION. HOSPITAL AND MEDICAL RECORDS HAVE BEEN REQUESTED BUT NOT YET PROVIDED

Description of Event or Problem · 0

PATIENT ALLEGEDLY RECEIVED AN IMPLANT ON (B)(6)2010 DUE TO DEEP VEIN THROMBOSIS (DVT)/PULMONARY EMBOLISM (PE). THE PATIENT ALLEGES TILT AND VENA CAVA PERFORATION. THE PATIENT FURTHER ALLEGES PAIN AND DISCOMFORT. (B)(6)2020, PER A REPORT FROM COMPUTED TOMOGRAPHY; ¿IMPRESSION: 1. IVC FILTER NOTED IN PLACE. THERE IS PENETRATION OF THE IVC WALL FROM THE METALLIC TINES AND CONTACT OF THE ANTERIOR VERTEBRAL COLUMN. NO BONY PENETRATION. THE PROXIMAL ASPECT OF THE IVC FILTER MAY POSSIBLY PERFORATE THE ANTERIOR WALL OF THE IVC BY ROUGHLY 3 MM. 2. THERE ARE 17 DEGREES OF ANTERIOR TILT OF THE IVC FILTER.¿

Additional Manufacturer Narrative · 1

(B)(4). EXEMPTION NUMBER E2016032. (B)(4). IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING 'BLOOD CLOTS, UNABLE TO BE RETRIEVED'. COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. FILTER RETRIEVAL IS OCCASIONALLY DIFFICULT. THIS IS WELL-KNOWN FROM PUBLISHED SCIENTIFIC LITERATURE WHERE FILTER RETRIEVALS ARE REFERRED TO AS SIMPLE VS. COMPLEX. SEVERAL CASE REPORTS PUBLISHED IN SCIENTIFIC LITERATURE DESCRIBE COMPLEX CASES WITH SUCCESSFUL ENDOVASCULAR FILTER RETRIEVALS USING ADDITIONAL, ADVANCED TECHNIQUES. UNKNOWN IF THE REPORTED PERIPHERAL VEIN THROMBOSIS IS DIRECTLY RELATED TO THE FILTER AND UNABLE TO IDENTIFY A CORRESPONDING FAILURE MODE AT THIS TIME. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Additional Manufacturer Narrative · 1

EXEMPTION NUMBER E2016032. (B)(4). IT HAS NOT BEEN POSSIBLE TO FURTHER INVESTIGATE OR EVALUATE THIS ALLEGED EVENT BASED ON THE LIMITED INFORMATION PROVIDED TO DATE VIA THE OPERATIVE NOTE STATING ¿CELECT FILTER IMPLANTED". COOK WILL REOPEN ITS INVESTIGATION IF FURTHER INFORMATION IS RECEIVED WARRANTING SUPPLEMENTATION IN ACCORDANCE WITH 21 C.F.R. 803.56. NO EVIDENCE TO SUGGEST THAT THIS DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS AND NOTHING INDICATES THAT THE FILTER DID NOT PERFORM AS INTENDED, E.G. INTENDED FOR THE PREVENTION OF RECURRENT PULMONARY EMBOLISM (PE) VIA PLACEMENT IN THE VENA CAVA.

Additional Manufacturer Narrative · 1

MANUFACTURER REFERENCE # (B)(4). EXEMPTION NUMBER E2016032. (B)(4). NAME AND ADDRESS FOR IMPORTER SITE: (B)(4). CORRECTED DATA BASED ON NEW INFORMATION RECEIVED: ADVERSE EVENT TO PRODUCT PROBLEM, SERIOUS INJURY TO MALFUNCTION. THE EVENT IS CURRENTLY UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE PROVIDED UPON CONCLUSION.

Description of Event or Problem · 1

THIS ADDITIONAL INFORMATION RECEIVED ON 11/14/2017 AS FOLLOWS: PATIENT RECEIVED AN IMPLANT ON (B)(6) 2010 VIA THE RIGHT COMMON FEMORAL VEIN DUE TO BLOOD CLOTS. PATIENT IS ALLEGING BLOOD CLOTS AND THAT THE DEVICE IS UNABLE TO BE RETRIEVED. NO DOCUMENTATION REGARDING ATTEMPTED RETRIEVAL WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736696 COOK CELECT FEMORAL VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR DTK WILLIAM COOK EUROPE

Patients

Seq Age Sex Outcome Treatment
1 Male Life Threatening