FDA Adverse Event Malfunction Summary report: N

RESTORIS PST MAKO OFFSET SHELL IMPACTOR

MDR report key: 6954554 · Received October 17, 2017

Report

Report Number
3005985723-2017-00500
Event Type
Malfunction
Date Received
October 17, 2017
Date of Event
September 19, 2017
Report Date
November 3, 2017
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K170593
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT: THE EVENT REPORTED THAT DURING IMPACTION IN A MAKO THA CASE, THE DOCTOR WAS IMPACTING THE CUP AND THE THREADS ON THE IMPACTION HANDLE STRIPED OUT AND THE IMPACTION HANDLE CAME UN-ATTACHED FROM THE CUP. THE CUP WAS IN ITS FINAL POSITION WHEN THIS HAPPENED. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION SHOWS THREAD DAMAGE. THREADS WERE BENT AND WORN AWAY IN A SINGLE AREA. SEE ATTACHED FIGURES. FUNCTIONAL INSPECTION CONFIRMED THAT A CUP WOULD NOT THREAD ONTO THE DEVICE. DEVICE HISTORY REVIEW: REVIEW OF DEVICE HISTORY RECORDS INDICATE THAT (B)(4) DEVICES WERE ACCEPTED INTO FINAL STOCK ON 07/27/2016 WITH NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: REVIEW OF COMPLAINTS FOR P/N 212260, L/N 32990 WITHIN THE TRACKWISE DATABASE SHOW NO OTHER COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSION: THE FAILURE WAS CONFIRMED. CORRECTIVE ACTION / PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

DURING A MAKO THA, THE DOCTOR WAS IMPACTING THE CUP. DURING IMPACTION THE THREADS ON THE IMPACTION HANDLE STRIPED OUT AND THE IMPACTION HANDLE CAME UN ATTACHED FROM THE CUP. THE CUP WAS IN ITS FINAL POSITION WHEN THIS HAPPENED.

Description of Event or Problem · 1

DURING A MAKO THA, THE DOCTOR WAS IMPACTING THE CUP. DURING IMPACTION THE THREADS ON THE IMPACTION HANDLE STRIPED OUT AND THE IMPACTION HANDLE CAME UN ATTACHED FROM THE CUP. THE CUP WAS IN ITS FINAL POSITION WHEN THIS HAPPENED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736068 RESTORIS PST MAKO OFFSET SHELL IMPACTOR STEREOTAXIC DEVICE, ROBOTICS OLO MAKO SURGICAL CORP. 32990

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization