FDA Adverse Event Malfunction Summary report: N

TRULIGHT

MDR report key: 6954213 · Received October 17, 2017

Report

Report Number
9681407-2017-00034
Event Type
Malfunction
Date Received
October 17, 2017
Date of Event
September 19, 2017
Report Date
September 19, 2017
Manufacturer
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Product Code
FSY
PMA / PMN Number
K102758
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IT IS KNOWN THAT AN IMPROPER INSTALLATION, MAINTENANCE OR A COLLISION CAN LEAD TO THE SPRING ARM COVER FALLING OFF. THE DEVICE WAS EVALUATED AND REPAIRED BY A TRUMPF MEDICAL SERVICE TECHNICIAN.

Description of Event or Problem · 1

THE FRONT COVER OF A SPRING ARM ON A TRUMPF MEDICAL SURGICAL LIGHT SYSTEM FELL OFF DURING A PROCEDURE AND INTO THE STERILE FIELD. THE COVER DID NOT CONTACT THE PATIENT AND NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
734252 TRULIGHT SURGICAL LIGHT FSY TRUMPF MEDIZIN SYSTEME GMBH + CO. KG 1574759

Patients

Seq Age Sex Outcome Treatment
1