S3 GAS BLENDER SYSTEM
Report
- Report Number
- 9611109-2017-00819
- Event Type
- Malfunction
- Date Received
- October 16, 2017
- Date of Event
- September 20, 2017
- Report Date
- December 7, 2017
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DTX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE DEVICE HAS BEEN SENT TO LIVANOVA (B)(4) FOR FURTHER EVALUATION. DURING THE INTERNAL INVESTIGATION, THE REPORTED FAILURE COULD NOT BE REPRODUCED. THE DEVICE UNDERWENT A ROUTINE MAINTENANCE AND A VALUE ADJUSTMENT AND IT WAS THEN RETURNED TO THE CUSTOMER BY LIVANOVA (B)(4). AS THE ISSUE COULD NOT BE REPRODUCED OR CONFIRMED, A ROOT CAUSE WAS NOT IDENTIFIED.
PATIENT INFORMATION WAS NOT PROVIDED. THE S3 GAS BLENDER 25-40-00 IS NOT DISTRIBUTED IN THE USA, BUT IT IS SIMILAR TO S5 GAS BLENDER SYSTEM 25-40-45, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K101046). LIVANOVA (B)(4) MANUFACTURES THE S3 GAS BLENDER SYSTEM. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE. THE SERVICE REPRESENTATIVE WAS UNABLE TO CONFIRM THE REPORTED ISSUE. THE DEVICE WAS TESTED WITHOUT A FAILURE. THE DEVICE HAS BEEN SENT TO LIVANOVA (B)(4) FOR FURTHER EVALUATION. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
LIVANOVA (B)(4) RECEIVED A REPORT THAT THE O2 SUPPLY OF THE S3 GAS BLENDER SYSTEM DROPPED DURING A PROCEDURE. THE TECHNICIAN USED AN O2 BOTTLE TO CONTINUE OXYGEN FLOW AND PREVENT PATIENT INJURY. THERE WAS NO REPORT OF PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 732744 | S3 GAS BLENDER SYSTEM | GAS CONTROL UNIT, CARDIOPULMONARY BYPASS | DTX | LIVANOVA DEUTSCHLAND | 25-40-00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |