FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 6953065 · Received October 16, 2017

Report

Report Number
9617229-2017-03294
Event Type
Injury
Date Received
October 16, 2017
Date of Event
September 22, 2017
Report Date
December 28, 2017
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
PMA / PMN Number
P990074
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE RELATED TO THE REPORTED EVENT OF DEFLATION WAS RECEIVED ON (B)(6) 2017 WITH LOT NUMBER 1831336 VISUAL ANALYSIS WAS PERFORMED WHICH IDENTIFIED FLAT CREASES, WHITE PARTICLES ON INNER SURFACE AND OUTER SURFACE OF DEVICE. LEAK TEST WAS PERFORMED WHICH IDENTIFIED NO LEAKAGE ON DEVICE. FILL INSPECTION WAS PERFORMED AND IDENTIFIED NO BLOCKAGE IN THE VALVE. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS NO ISSUES FOUND RELATED WITH THE MANUFACTURING PROCESS. NO WERE OBSERVED OPENINGS ON THE DEVICE OR ISSUES RELATED WITH THE COMPLAINT (DEFLATION).

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. DEVICE LABELING: POTENTIAL ADVERSE EVENTS THAT MAY OCCUR WITH SALINE-FILLED BREAST IMPLANT SURGERY INCLUDE: REOPERATION, PAIN, WRINKLING, ASYMMETRY, IMPLANT PALPABILITY/VISIBILITY, IMPLANT REMOVAL, CAPSULAR CONTRACTURE, CHANGES IN NIPPLE AND BREAST SENSATION, IMPLANT DISPLACEMENT/MIGRATION, IMPLANT DEFLATION, SCARRING, INFECTION, HEMATOMA/SEROMA, BREASTFEEDING COMPLICATIONS, IMPLANT EXTRUSION, NECROSIS, DELAYED WOUND HEALING, BREAST TISSUE ATROPHY/CHEST WALL DEFORMITY, CALCIUM DEPOSITS, AND LYMPHADENOPATHY. DEFLATION ¿ BREAST IMPLANTS ARE NOT LIFETIME DEVICES. SALINE BREAST IMPLANTS DEFLATE WHEN THE SHELL DEVELOPS A TEAR OR HOLE. DEFLATION CAN OCCUR AT ANY TIME AFTER IMPLANTATION, BUT THEY ARE MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING THINGS MAY CAUSE IMPLANTS TO DEFLATE: DAMAGE BY SURGICAL INSTRUMENTS; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G., DURING CLOSED CAPSULOTOMY, WHICH IS CONTRAINDICATED); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. LABORATORY STUDIES HAVE IDENTIFIED SOME OF THE CAUSES OF DEFLATION FOR ALLERGAN¿S PRODUCT; HOWEVER, IT IS NOT CONCLUSIVELY KNOWN WHETHER THESE TESTS HAVE IDENTIFIED ALL CAUSES OF DEFLATION.

Description of Event or Problem · 1

THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 1

THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 1

PATIENT REPORTED RIGHT SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
730553 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 1831336

Patients

Seq Age Sex Outcome Treatment
1 26 YR Required Intervention