FDA Adverse Event Malfunction Summary report: N

EVIS EXERA II DUODENOVIDEOSCOPE

MDR report key: 6947142 · Received October 13, 2017

Report

Report Number
8010047-2017-01461
Event Type
Malfunction
Date Received
October 13, 2017
Date of Event
September 16, 2017
Report Date
November 16, 2017
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDT
PMA / PMN Number
PK143153
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE DEVICE EVALUATION RESULT. THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. BUT WAS RETURNED TO OLYMPUS (B)(4). FOLLOWING ADDITIONAL HIGH LEVEL DISINFECTION AT (B)(4), THE SUBJECT DEVICE WAS SENT TO A THIRD PARTY LABORATORY FOR ADDITIONAL MICROBIOLOGICAL TESTING. IN THE ADDITIONAL TEST, THE TESTING INDICATED NO MICROBIAL GROWTH FOR THE SUBJECT DEVICE.

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC. OMSC REVIEWED THE MANUFACTURING HISTORY OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT DURING ROUTINE SURVEILLANCE CULTURING TEST BY THE FACILITY, THE SUBJECT DEVICE TESTED POSITIVE FOR UNSPECIFIED BACTERIA. THE CHANNELS OF THE SUBJECT DEVICE TESTED POSITIVE FOR FOLLOWING CFU OF BACTERIA. THE INSTRUMENT CHANNEL: 19 CFU / 100ML, THE SUCTION CHANNEL: 6 CFU / 100ML, THE AIR/WATER CHANNEL: 9 CFU / 100ML. THE SUBJECT DEVICE HAD BEEN REPROCESSED USING AN OLYMPUS AUTOMATED ENDOSCOPE REPROCESSR MODEL ETD3 (NOT AVAILABLE IN THE U.S.) WITH PERACETIC ACID. THE FORCEPS ELEVATOR WAS REPLACED IN (B)(6) 2017 ACCORDING TO AN OLYMPUS FIELD CORRECTIVE ACTION. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
729788 EVIS EXERA II DUODENOVIDEOSCOPE DUODENOVIDEOSCOPE FDT OLYMPUS MEDICAL SYSTEMS CORP. TJF-Q180V

Patients

Seq Age Sex Outcome Treatment
1