FDA Adverse Event Injury Summary report: N

BIOMET 360 TIBIAL TRAY WITH TI LOCKING BAR & SCREW

MDR report key: 6943837 · Received October 12, 2017

Report

Report Number
0001825034-2017-08489
Event Type
Injury
Date Received
October 12, 2017
Date of Event
June 6, 2017
Report Date
October 12, 2017
Manufacturer
ZIMMER BIOMET, INC.
Product Code
JWH
PMA / PMN Number
PK093293
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). CONCOMITANT MEDICAL PRODUCTS: BMT SPLINED KNEE STM V2 18X40, CAT#: 148293, LOT#: 818780; BMT 360 TIB AUG 83X5MM, CAT#: 185226, LOT#: 059760; BMT 360 TIB AUG 83X5MM, CAT#: 185226, LOT#: 304860; BMT 360 TIB LG CRUCIATE WING, CAT#: 185651, LOT#: 790440; BMT 360 TIB 5.0 OFFSET ADAPTER, CAT#: 185211, LOT#: 874310; E1 VNGD PS+ TIB BRG 79/83X16, CAT#: EP-183766, LOT#: 111750; VAN PS OPEN INTL FEM-LT 70 CAT#. 183132, LOT#. 442030. REPORTED EVENT WAS CONFIRMED BY REVIEW OF X-RAYS. DHR WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. REVIEW OF THE COMPLAINT HISTORY DETERMINED THAT NO FURTHER ACTION IS REQUIRED AS NO WERE TRENDS IDENTIFIED. ROOT CAUSE WAS UNABLE TO BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. THE RISK FOR THE REPORTED ISSUE IS ADDRESSED IN I/O RISK TABLE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034 - 2017 - 04596, 0001825034 - 2017 - 08486, 0001825034 - 2017 - 08487, 0001825034 - 2017 - 08488, 0001825034 - 2017 - 08494, 0001825034 - 2017 - 08495.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS REVISED TO ADDRESS FEMORAL COMPONENT LOOSENING. FURTHER REVIEW OF THE PROVIDED PATIENT'S X-RAYS DETERMINED THAT RADIOLUCENCY IS ALSO PRESENT AT TIBIAL IMPLANT CEMENT-BONE AND METAL-BONE INTERFACES CONSISTENT WITH LOOSENING. ADDITIONALLY, THE TIBIAL TRAY IS NOT PERPENDICULAR TO THE LONG AXIS OF THE TIBIA ON THE AP VIEW AND RATHER EXHIBITS SEVERAL DEGREES TILT WITH SUGGESTION OF SUBSIDENCE OF THE MEDIAL TIBIAL TRAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723307 BIOMET 360 TIBIAL TRAY WITH TI LOCKING BAR & SCREW PROSTHESIS, KNEE JWH ZIMMER BIOMET, INC. N/A 952960

Patients

Seq Age Sex Outcome Treatment
1 53 YR Hospitalization| R