FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE 3/10 ML/CC 31 G X 6 MM HALF UNIT

MDR report key: 6942473 · Received October 11, 2017

Report

Report Number
1920898-2017-00270
Event Type
Malfunction
Date Received
October 11, 2017
Date of Event
September 18, 2017
Report Date
November 30, 2017
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
K024112
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: A SAMPLE IS NOT AVAILABLE FOR EVALUATION. PRODUCT DESCRIPTION: SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAG, DATE(S) PACKAGING: 25APR2017 THRU 27APR2017, MACHINE(S) PACKAGED ON: (B)(4). DEVICE HISTORY RECORD REVIEW ¿ A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 7065710 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. ASSEMBLED SYRINGES ¿ THERE WAS ONE (1) BATCH OF MATERIAL # 700003988 (SYRINGE 0.3ML ASM 31GA 6MM TW (B)(4)) THAT WENT INTO THE PACKAGED BATCH# 7065710. BATCH# 7086920, DATE(S) OF MANUFACTURE: 22APR2017 THRU 26APR2017, MACHINE(S) MANUFACTURED ON: (B)(4). PRINTED BARRELS ¿ THERE WERE TWO (2) BATCH OF MATERIAL# 700003987 (BARREL 0.3ML PRINTED SM700176) THAT WENT INTO PACKAGED BATCH# 7065710. BATCH# 7103598, DATE(S) OF MANUFACTURE: 22APR2017 THRU 29APR2017, MACHINE(S) MANUFACTURED ON: (B)(4). BATCH# 7093853, DATE(S) OF MANUFACTURE: 14APR2017 THRU 18APR2017, MACHINE(S) MANUFACTURED ON: (B)(4). THERE WERE THREE (3) NOTIFICATIONS [(B)(4)] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. CONCLUSION: WITHOUT A SAMPLE, AN ABSOLUTE ROOT CAUSE FOR THIS INCIDENT CANNOT BE DETERMINED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD INSULIN SYRINGE 3/10 ML/CC 31 G X 6 MM HALF UNIT CAME WITH THE STOPPER AT AN ANGLE AND CANNOT DRAW THE ACCURATE DOSE, WHILE THE SCALE MARKING PRINT IS OFF. FOUND BEFORE USE. NO SERIOUS INJURY OR MEDICAL INTERVENTION NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
718981 BD INSULIN SYRINGE 3/10 ML/CC 31 G X 6 MM HALF UNIT PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 7065710

Patients

Seq Age Sex Outcome Treatment
1 Other