FDA Adverse Event
Other
Summary report: N
UNO
MDR report key: 694243
·
Received March 29, 2006
Report
- Report Number
- 1225750-2006-00004
- Event Type
- Other
- Date Received
- March 29, 2006
- Date of Event
- March 19, 2006
- Report Date
- March 29, 2006
- Manufacturer
- LIKO AB
- Product Code
- FSA
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- PA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
ON MARCH 19, 2006 WHILE TRANSFERRING A RESIDENT USING AN UNO 102 PT LIFT, THE FACILITY REPORTED THAT WITH 3 OF THE 4 LIFT STRAPS IN PLACE, THE LIFT ROSE QUICKLY WITHOUT USE IF THE HAND CONTROL. THE RESIDENT FELL OUT OF THE SLING SUFFERING A LACERATION TO THE HEAD. THREE DAYS LATER A LIKO REP. WAS ALLOWED TO VIEW AND INSPECT THE EQUIPMENT. IT WAS FOUND THAT THE SAFETY LATCHES ON THE SLINGBAR WERE MISSING AND THE LIFT DID NOT HAVE APPEARANCE OF BEING PROPERLY MAINTAINED. THE HAND CONTROL IS BEING RETURNED TO LIKO INC FOR ANALYSIS, AND A FOLLOW UP REPORT WILL BE FORWARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNO | PATIENT LIFT/HOIST | FSA | LIKO AB | UNO 102EE | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 93 YR | Other |