FDA Adverse Event Malfunction Summary report: N

ARC INSTATEMP

MDR report key: 6940411 · Received October 11, 2017

Report

Report Number
3011197139-2017-00108
Event Type
Malfunction
Date Received
October 11, 2017
Date of Event
September 11, 2017
Report Date
October 5, 2017
Manufacturer
ARC DEVICES USA INC.
Product Code
FLL
PMA / PMN Number
K152905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

CUSTOMER STATES THE THERMOMETER WAS FOUND EXPLODED IN THE BATHROOM CLOSET WHERE THE UNIT WAS SITTING UNATTENDED. NO ONE WAS HURT. ARC REQUESTED THE CUSTOMER TO RETURN THE PRODUCT. A LAB ANALYSIS FOR THIS TYPE OF FAILURE WAS PERFORMED ON SIMILAR DEVICE. THE ANALYSIS OF THE BATTERIES INCLUDED AN OPTICAL MICROSCOPY, CT SCAN AND 2D X-RAY AND INDICATED THE FAILURE WAS DUE TO A SINGLE COMPONENT (BATTERY) MALFUNCTION. REPORT, ATTACHED. THE SUSPECT BATTERIES HAVE NOT BEEN USED IN PRODUCTION SINCE 17 JULY 2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
721109 ARC INSTATEMP THERMOMETER, ELECTRONIC, CLINICAL FLL ARC DEVICES USA INC. CF1-0196

Patients

Seq Age Sex Outcome Treatment
1