QUINTON Q-TEL RMS
Report
- Report Number
- MW5072668
- Event Type
- Malfunction
- Date Received
- October 10, 2017
- Date of Event
- October 4, 2017
- Report Date
- October 5, 2017
- Manufacturer
- MORTARA INSTRUMENT, INC.
- Product Code
- MHX
- UDI-DI
- 00312345029604
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DURING A MONITORED CARDIAC REHABILITATION EXERCISE SESSION, OUR G-TEL RMS ECG MONITORING SYSTEM CRASHED. FIVE PTS WERE ON MONITORS AT THE TIME OF THE CRASH. THE SYSTEM DID NOT AUTOMATICALLY REBOOT, AND IT DID NOT RECOVER SESSION DATA. THE SYSTEM HAD TO BE MANUALLY RESTARTED, AND EACH PT INDIVIDUALLY RE-ADMITTED. ALL OF THE DATA FOR EACH PT'S SESSION UP TO THE TIME OF THE CRASH WAS LOST. ECG DATA, BLOOD PRESSURES, RPE, SPO2, AND EXERCISE MODALITIES AND WORKLOADS WERE LOST. THIS PRESENTS A SERIOUS ISSUE AS CARDIAC EVENTS COULD HAVE GONE UNDETECTED DURING THE TIME THE SYSTEM WAS BEING RESTARTED AND THE PTS WERE BEING RE-ADMITTED. ADDITIONALLY, ANY SIGNIFICANT EVENTS THAT OCCURRED DURING THE SESSION PRIOR TO THE CRASH CANNOT BE DOCUMENTED AND RELAYED TO THE CARDIOLOGIST. THE COMPANY HAS BEEN NOTIFIED OF THIS EVENT, AND THEY STATED THAT THEY HAVE BEEN AWARE OF THEIR SYSTEM CRASHING FOR "A FEW MONTHS NOW" AND THAT THEY ARE "WORKING ON A SOLUTION." OUR DEPARTMENT WAS NOT NOTIFIED OF THE ISSUE PRIOR TO YESTERDAY'S OCCURRENCE, AND WE WERE UNAWARE THAT THE SYSTEM WE ARE UTILIZING HAS HAD THIS KNOWN DEFECT. THIS DEFECT SIGNIFICANTLY AFFECTS PT CARE IN A NEGATIVE MANNER AND CAN POTENTIALLY LEAD TO SERIOUS CARDIAC EVENTS GOING UNDETECTED.
DURING A MONITORED CARDIAC REHABILITATION EXERCISE SESSION, OUR G-TEL RMS ECG MONITORING SYSTEM CRASHED. FIVE PTS WERE ON MONITORS AT THE TIME OF THE CRASH. THE SYSTEM DID NOT AUTOMATICALLY REBOOT, AND IT DID NOT RECOVER SESSION DATA. THE SYSTEM HAD TO BE MANUALLY RESTARTED, AND EACH PT INDIVIDUALLY RE-ADMITTED. ALL OF THE DATA FOR EACH PT'S SESSION UP TO THE TIME OF THE CRASH WAS LOST. ECG DATA, BLOOD PRESSURES, RPE, SPO2, AND EXERCISE MODALITIES AND WORKLOADS WERE LOST. THIS PRESENTS A SERIOUS ISSUE AS CARDIAC EVENTS COULD HAVE GONE UNDETECTED DURING THE TIME THE SYSTEM WAS BEING RESTARTED AND THE PTS WERE BEING RE-ADMITTED. ADDITIONALLY, ANY SIGNIFICANT EVENTS THAT OCCURRED DURING THE SESSION PRIOR TO THE CRASH CANNOT BE DOCUMENTED AND RELAYED TO THE CARDIOLOGIST. THE COMPANY HAS BEEN NOTIFIED OF THIS EVENT, AND THEY STATED THAT THEY HAVE BEEN AWARE OF THEIR SYSTEM CRASHING FOR "A FEW MONTHS NOW" AND THAT THEY ARE "WORKING ON A SOLUTION." OUR DEPARTMENT WAS NOT NOTIFIED OF THE ISSUE PRIOR TO YESTERDAY'S OCCURRENCE, AND WE WERE UNAWARE THAT THE SYSTEM WE ARE UTILIZING HAS HAD THIS KNOWN DEFECT. THIS DEFECT SIGNIFICANTLY AFFECTS PT CARE IN A NEGATIVE MANNER AND CAN POTENTIALLY LEAD TO SERIOUS CARDIAC EVENTS GOING UNDETECTED.
DURING A MONITORED CARDIAC REHABILITATION EXERCISE SESSION, OUR G-TEL RMS ECG MONITORING SYSTEM CRASHED. FIVE PTS WERE ON MONITORS AT THE TIME OF THE CRASH. THE SYSTEM DID NOT AUTOMATICALLY REBOOT, AND IT DID NOT RECOVER SESSION DATA. THE SYSTEM HAD TO BE MANUALLY RESTARTED, AND EACH PT INDIVIDUALLY RE-ADMITTED. ALL OF THE DATA FOR EACH PT'S SESSION UP TO THE TIME OF THE CRASH WAS LOST. ECG DATA, BLOOD PRESSURES, RPE, SPO2, AND EXERCISE MODALITIES AND WORKLOADS WERE LOST. THIS PRESENTS A SERIOUS ISSUE AS CARDIAC EVENTS COULD HAVE GONE UNDETECTED DURING THE TIME THE SYSTEM WAS BEING RESTARTED AND THE PTS WERE BEING RE-ADMITTED. ADDITIONALLY, ANY SIGNIFICANT EVENTS THAT OCCURRED DURING THE SESSION PRIOR TO THE CRASH CANNOT BE DOCUMENTED AND RELAYED TO THE CARDIOLOGIST. THE COMPANY HAS BEEN NOTIFIED OF THIS EVENT, AND THEY STATED THAT THEY HAVE BEEN AWARE OF THEIR SYSTEM CRASHING FOR "A FEW MONTHS NOW" AND THAT THEY ARE "WORKING ON A SOLUTION." OUR DEPARTMENT WAS NOT NOTIFIED OF THE ISSUE PRIOR TO YESTERDAY'S OCCURRENCE, AND WE WERE UNAWARE THAT THE SYSTEM WE ARE UTILIZING HAS HAD THIS KNOWN DEFECT. THIS DEFECT SIGNIFICANTLY AFFECTS PT CARE IN A NEGATIVE MANNER AND CAN POTENTIALLY LEAD TO SERIOUS CARDIAC EVENTS GOING UNDETECTED.
DURING A MONITORED CARDIAC REHABILITATION EXERCISE SESSION, OUR G-TEL RMS ECG MONITORING SYSTEM CRASHED. FIVE PTS WERE ON MONITORS AT THE TIME OF THE CRASH. THE SYSTEM DID NOT AUTOMATICALLY REBOOT, AND IT DID NOT RECOVER SESSION DATA. THE SYSTEM HAD TO BE MANUALLY RESTARTED, AND EACH PT INDIVIDUALLY RE-ADMITTED. ALL OF THE DATA FOR EACH PT'S SESSION UP TO THE TIME OF THE CRASH WAS LOST. ECG DATA, BLOOD PRESSURES, RPE, SPO2, AND EXERCISE MODALITIES AND WORKLOADS WERE LOST. THIS PRESENTS A SERIOUS ISSUE AS CARDIAC EVENTS COULD HAVE GONE UNDETECTED DURING THE TIME THE SYSTEM WAS BEING RESTARTED AND THE PTS WERE BEING RE-ADMITTED. ADDITIONALLY, ANY SIGNIFICANT EVENTS THAT OCCURRED DURING THE SESSION PRIOR TO THE CRASH CANNOT BE DOCUMENTED AND RELAYED TO THE CARDIOLOGIST. THE COMPANY HAS BEEN NOTIFIED OF THIS EVENT, AND THEY STATED THAT THEY HAVE BEEN AWARE OF THEIR SYSTEM CRASHING FOR "A FEW MONTHS NOW" AND THAT THEY ARE "WORKING ON A SOLUTION." OUR DEPARTMENT WAS NOT NOTIFIED OF THE ISSUE PRIOR TO YESTERDAY'S OCCURRENCE, AND WE WERE UNAWARE THAT THE SYSTEM WE ARE UTILIZING HAS HAD THIS KNOWN DEFECT. THIS DEFECT SIGNIFICANTLY AFFECTS PT CARE IN A NEGATIVE MANNER AND CAN POTENTIALLY LEAD TO SERIOUS CARDIAC EVENTS GOING UNDETECTED.
DURING A MONITORED CARDIAC REHABILITATION EXERCISE SESSION, OUR G-TEL RMS ECG MONITORING SYSTEM CRASHED. FIVE PTS WERE ON MONITORS AT THE TIME OF THE CRASH. THE SYSTEM DID NOT AUTOMATICALLY REBOOT, AND IT DID NOT RECOVER SESSION DATA. THE SYSTEM HAD TO BE MANUALLY RESTARTED, AND EACH PT INDIVIDUALLY RE-ADMITTED. ALL OF THE DATA FOR EACH PT'S SESSION UP TO THE TIME OF THE CRASH WAS LOST. ECG DATA, BLOOD PRESSURES, RPE, SPO2, AND EXERCISE MODALITIES AND WORKLOADS WERE LOST. THIS PRESENTS A SERIOUS ISSUE AS CARDIAC EVENTS COULD HAVE GONE UNDETECTED DURING THE TIME THE SYSTEM WAS BEING RESTARTED AND THE PTS WERE BEING RE-ADMITTED. ADDITIONALLY, ANY SIGNIFICANT EVENTS THAT OCCURRED DURING THE SESSION PRIOR TO THE CRASH CANNOT BE DOCUMENTED AND RELAYED TO THE CARDIOLOGIST. THE COMPANY HAS BEEN NOTIFIED OF THIS EVENT, AND THEY STATED THAT THEY HAVE BEEN AWARE OF THEIR SYSTEM CRASHING FOR "A FEW MONTHS NOW" AND THAT THEY ARE "WORKING ON A SOLUTION." OUR DEPARTMENT WAS NOT NOTIFIED OF THE ISSUE PRIOR TO YESTERDAY'S OCCURRENCE, AND WE WERE UNAWARE THAT THE SYSTEM WE ARE UTILIZING HAS HAD THIS KNOWN DEFECT. THIS DEFECT SIGNIFICANTLY AFFECTS PT CARE IN A NEGATIVE MANNER AND CAN POTENTIALLY LEAD TO SERIOUS CARDIAC EVENTS GOING UNDETECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 718899 | QUINTON Q-TEL RMS | CARDIAC REHABILITATION SYSTEM | MHX | MORTARA INSTRUMENT, INC. | VERSION 4.1.0 | 00312345029604 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other | ||
| 2 | 54 YR | Other | ||
| 3 | 86 YR | Other | ||
| 4 | 70 YR | Other | ||
| 5 | 79 YR | Other |