FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 693476 · Received March 29, 2006

Report

Report Number
2243072-2006-00129
Event Type
Other
Date Received
March 29, 2006
Date of Event
March 3, 2006
Report Date
March 27, 2006
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT READING OF 23 FOLLOWED BY A READING OF 220. FELT MORE LIKE TO THE 220 AND THINKS THE METER IS READING LOW. ANALYSIS RUN CONSENSUS ERROR GRID USING MEAN READING (122) AS REF POINT VS BD READINGS (23,220) YIELDED SOME DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 5064111

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other