FDA Adverse Event
Other
Summary report: N
BD PARADIGM LINK BLOOD GLUCOSE MONITOR
MDR report key: 693476
·
Received March 29, 2006
Report
- Report Number
- 2243072-2006-00129
- Event Type
- Other
- Date Received
- March 29, 2006
- Date of Event
- March 3, 2006
- Report Date
- March 27, 2006
- Manufacturer
- BECTON DICKINSON
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
CONSUMER CALLED TO REPORT READING OF 23 FOLLOWED BY A READING OF 220. FELT MORE LIKE TO THE 220 AND THINKS THE METER IS READING LOW. ANALYSIS RUN CONSENSUS ERROR GRID USING MEAN READING (122) AS REF POINT VS BD READINGS (23,220) YIELDED SOME DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD PARADIGM LINK BLOOD GLUCOSE MONITOR | BD TEST STRIPS | NBW | BECTON DICKINSON | NA | 5064111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |