MODULAR EBS KNEE JOINT
Report
- Report Number
- 9616494-2017-00045
- Event Type
- Malfunction
- Date Received
- October 10, 2017
- Report Date
- October 10, 2017
- Manufacturer
- OTTO BOCK HEALTHCARE GMBH
- Product Code
- ISY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE COMPLAINT NOTIFICATION ASSOCIATED WITH THIS DEVICE DESCRIBED THAT THE KNEE JOINT IS BLOCKED. THE USER DID NOT FALL, NO SERIOUS INJURIES WERE SUSTAINED AS A RESULT OF THE FAILURE AND NO MEDICAL TREATMENT WAS REQUIRED. THE EVALUATION OF THIS SPECIFIC DEVICE WAS CONDUCTED AT THE MANUFACTURER'S AFTER SALES SERVICE CENTER ON (B)(6) 2017. AFTER EVALUATION WE CAN CONFIRM THAT ONE BOLT IN THE UPPER POSTERIOR SECTION OF THE KNEE JOINT WAS LOOSE, PREVENTING THE KNEE JOINT FROM BENDING. AN EXACT REASON FOR THE LOSS/LOOSENING OF THE BOLT COULD NOT BE DETERMINED. IN THIS SPECIFIC CASE THE FAILURE DID NOT LEAD TO A FALL AND/OR SERIOUS INJURY, BUT IT COULD CAUSE OR CONTRIBUTE TO A SERIOUS INJURY IF IT WERE TO RECUR. THEREFORE, WE REPORT THIS FAILURE ACCORDING TO GUIDANCE FOR INDUSTRY AND FDA STAFF ON "MDR FOR MANUFACTURERS" CHAPTER 2.15.
KNEE JOINT WAS SENT IN FOR REPAIR. CUSTOMER STATED THAT THE KNEE JOINT IS BLOCKED. THE USER DID NOT FALL AND SUSTAINED NO SERIOUS INJURIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 714281 | MODULAR EBS KNEE JOINT | MODULAR EBS KNEE JOINT | ISY | OTTO BOCK HEALTHCARE GMBH | 3R60 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |