I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2017-00455
- Event Type
- Malfunction
- Date Received
- October 9, 2017
- Date of Event
- February 1, 2016
- Report Date
- October 9, 2017
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749001917
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT # (B)(4). APOC LABELING WAS EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT. THE INVESTIGATION WAS COMPLETED ON 04/22/2016. RETAIN PRODUCT WAS TESTED AND FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE. INVESTIGATION: THE DEVICE HISTORY RECORD (DHR) FOR THIS LOT WAS REVIEWED. THE LOT PASSED FINISHED GOODS (FG) RELEASE CRITERIA. FORTY RETAINED CARTRIDGES FROM THE LOT WERE TESTED USING WHOLE BLOOD (WB) AND I-STAT LEVEL 1 CONTROL (L1). TESTING MET THE ACCEPTANCE CRITERIA FOUND IN Q04.01.003 REV. W, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED. ASSESSMENT: THE COMPLAINT INVESTIGATION CONCLUDED THERE WAS NO PRODUCT DEFICIENCY AND THAT THE I-STAT CARTRIDGE LOT IS MEETING SPECIFICATION. THERE ARE NO REPEATS ON THE I-STAT ON THE SAME DAY. A DIFFERENT SAMPLE WAS DRAWN AND SENT TO THE LABORATORY.
ON (B)(6) 2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT RESULTS ON A (B)(6) YEAR OLD FEMALE PATIENT IN THE ER WITH NAUSEA, VOMITING AND DEHYDRATION. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. RESULTS (B)(6) 2016: I-STAT: VISTA: NA+ 110, 143. K+ 2.8, 3.7. CL- >140, 108. I-CA+ 0.67, ----. BUN 28, 17. HCT 29%, 39.8%. HGB 9.9, 13.1. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY APOC, HOWEVER THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED MAY 8TH TO 12TH, 2017 AT ABBOTT POINT OF CARE (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 709850 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H15309 | 10054749001917 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR |