FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 6928006 · Received October 9, 2017

Report

Report Number
1823260-2017-02217
Event Type
Malfunction
Date Received
October 9, 2017
Date of Event
September 19, 2017
Report Date
October 9, 2017
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE CUSTOMER RECEIVED QUESTIONABLE RESULTS FROM COAGUCHEK XS SERIAL NUMBER (B)(4). THE INITIAL RESULT AT 9:15 AM WAS 4.1 INR AND THE REPEAT RESULT WAS 2.9 INR. ANOTHER TEST PERFORMED AT 11:00 AM WITH TEST STRIP LOT 22385523 HAD A RESULT OF 4.3 INR. THE CUSTOMER REPORTED ALL OF THE RESULTS TO THE DOCTOR AND NONE WERE TRUSTED DUE TO DIFFERENCE. A VENIPUNCTURE WAS THEN PERFORMED AT 1:00 PM AND THE RESULT WAS 3.4 INR. THE REAGENT USED BY THE LABORATORY WAS NOT KNOWN. THE CUSTOMER WAS NOT TREATED BASED ON THE METER RESULTS. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT. THE CUSTOMER HAD NO KNOWN ANEMIA, POLYCYTHEMIA, HEPARINS, ANTIPHOSPHOLIPID ANTIBODIES, OR OTHER ANTICOAGULANTS. THE THERAPEUTIC RANGE WAS 2.0-3.0 INR. THE RETURNED METER AND STRIPS FROM LOT 22385523 WERE RECEIVED FOR INVESTIGATION. THEY WERE TESTED IN COMPARISON TO A RETENTION METER AND MASTERLOT STRIPS. TWO HUMAN BLOOD SAMPLES FROM WARFARIN DONORS AND INTERNAL REFERENCE METERS WERE USED. DONOR INR: 2.2 INR AND 2.5 INR . DONOR HCT: 52% AND 49%. TESTING RESULTS: DONOR 1: MASTER LOT / CUSTOMER STRIP AND METER, 2.3 INR/ 2.3 INR. DONOR 2: MASTER LOT / CUSTOMER STRIP AND METER, 2.5 INR/ 2.5 INR. ALL RESULTS WERE WITHIN THE SPECIFIED MAXIMUM DIFFERENCE BETWEEN MEASUREMENTS. NO ERROR MESSAGES OCCURRED. THE RETURNED AND THE RETENTION MATERIAL MET THE SPECIFICATION. NO INFORMATION WAS PROVIDED IN THE COMPLAINT CASE THAT WOULD POINT TO A CAUSE FOR THE RESULT DISCREPANCY. RELEVANT RETENTION TEST STRIPS (LOT 223855-23) WERE TESTED IN COMPARISON WITH THE CURRENT MASTER LOT COAGUCHEK XS PT TEST STRIP (LOT 124158-80). FOR THIS PURPOSE TWO HUMAN BLOOD SAMPLES FROM MARCUMAR DONORS AND TWO INTERNAL REFERENCE METERS WERE USED. NO ERROR MESSAGES OCCURRED. RETENTION SAMPLES WERE ACCEPTABLE AND RETENTION MATERIAL PERFORMED AS SPECIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
709920 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 19449123

Patients

Seq Age Sex Outcome Treatment
1 79 YR