I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2017-00409
- Event Type
- Malfunction
- Date Received
- October 7, 2017
- Date of Event
- June 18, 2016
- Report Date
- October 6, 2017
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- UDI-DI
- 10054749001917
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). APOC LABELING WAS EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT. THE INVESTIGATION WAS COMPLETED ON 10/31/2016. RETAIN PRODUCT WAS TESTED AND FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. INVESTIGATION: A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. Z (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). CUSTOMER COMPLAINT WAS NOT REPRODUCED. ASSESSMENT: THE COMPLAINT INVESTIGATION CONCLUDED THERE WAS NO PRODUCT DEFICIENCY AND THAT THE I-STAT CARTRIDGE LOT IS MEETING SPECIFICATION. NO REPEATS ON THE I-STAT OR LABORATORY ON THE SAME SAMPLE ON THE SAME DAY.
ON 07/13/2016, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT CHLORIDE (CL), AND SODIUM (NA) RESULT ON A (B)(6) YEAR OLD FEMALE PATIENT PRESENTED WITH HYPONATREMIA. THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. THERE WAS NO REPEAT ON I-STAT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: I-STAT; DATE: (B)(6) 2016; TIME: 10:59; NA: 115; CL: >140. DXC; UNK; UNK; 140; 105. AT THE TIME OF THE EVENT THERE WAS NO INDICATION OF A PRODUCT MALFUNCTION BASED ON THE INFORMATION PROVIDED AND ASSESSED BY APOC, HOWEVER THIS MDR IS A RETROSPECTIVE FILING IN RESPONSE TO AN OBSERVATION FROM AN FDA INSPECTION CONDUCTED MAY 8TH TO 12TH, 2017 AT ABBOTT POINT OF CARE ((B)(4)).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 706254 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | H16040 | 10054749001917 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |