FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE (SEE H10 FOR EC8+)

MDR report key: 6926811 · Received October 7, 2017

Report

Report Number
2245578-2017-00547
Event Type
Malfunction
Date Received
October 7, 2017
Date of Event
September 24, 2017
Report Date
November 30, 2017
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
UDI-DI
10054749001917
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). APOC LABELING WAS EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT. CARTRIDGE INFORMATION ASSOCIATED WITH THE CUSTOMER'S REPORTED EVENT PRODUCT NAME: I-STAT CARTRIDGE EC8+/ CATALOG # 03P79-25 / PRODUCT CODE: CHL; LOT# K17099 / LOT EXPIRATION: 10/28/2017 / MFG DATE: (B)(4) 2017/ 510K: K940918.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4).THE INVESTIGATION WAS COMPLETED ON 11/23/2017. RETAIN PRODUCT WAS TESTED AND FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE.

Description of Event or Problem · 1

ON (B)(6) 2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+, EC8+ CARTRIDGES THAT YIELDED SUSPECTED DISCREPANT RESULTS ON A PATIENT (DOG). CUSTOMER REPORTS TWO UNEXPECTED POTASSIUM RESULTS OF > 9 ON THE SAME SAMPLE. THE SAME SAMPLE WAS THEN RUN ON THE SITE'S CATALYST ANALYZER, K RESULT WAS 3.2. SITE COULD NOT RETEST AS THE ANIMAL WAS SCHEDULED TO BE EUTHANIZED, HOWEVER A SAMPLE FROM A DIFFERENT DOG WAS RUN ON BOTH CARTRIDGE TYPES AND GENERATED EXPECTED RESULTS. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE RESULT OF >9 RESULTED ON TWO DIFFERENT CARTRIDGE TYPES ON THE SAME SAMPLE. HEMOLYSIS IS SUSPECTED BUT NOT CONFIRMED AT THIS TIME. APOC HAS REQUESTED ALL OTHER TESTS FROM BOTH CARTRIDGES. THE INVESTIGATION IS UNDERWAY.

Description of Event or Problem · 1

NA.