FDA Adverse Event Malfunction Summary report: N

I-STAT DOWNLOADER, NETWORK

MDR report key: 6926795 · Received October 7, 2017

Report

Report Number
2245578-2017-00548
Event Type
Malfunction
Date Received
October 7, 2017
Date of Event
September 26, 2017
Report Date
December 19, 2017
Manufacturer
ABBOTT POINT OF CARE
Product Code
RHA
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 1

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 12/18/2017. THE CUSTOMER REPORTED THAT NETWORK DOWNLOADER (DRN) S/N (B)(4) BEGAN TO EMIT SMOKE WHEN AN ANALYZER WAS DOCKED. DRN S/N (B)(4) WAS RETURNED AS "TRANSFER" TO FLEX. BASED ON MATERIAL DESTRUCT ORDER MDO # OCT 2017-02, DRN S/N (B)(4) WAS SCRAPPED WITHOUT ANY INSPECTION OR EVALUATION PERFORMED. A ROCKETWARE SEARCH SPANNING SIX MONTHS REVEALED ONE POTENTIALLY RELATED INCIDENT AND NO EVIDENCE OF A TREND. NO DEFICIENCY HAS BEEN IDENTIFIED.

Description of Event or Problem · 1

ON (B)(6) 2017, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED VISUAL SMOKE COMING FROM NETWORK DOWNLOADER (B)(4). THE DEVICE WAS TAKEN OUT OF SERVICE, REPLACED AT NO CHARGE AND RETURNING FOR INVESTIGATION. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707345 I-STAT DOWNLOADER, NETWORK DOWNLOADER, NETWORK RHA ABBOTT POINT OF CARE NA NA

Patients

Seq Age Sex Outcome Treatment
1