FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - INDIGO¿ DRILL

MDR report key: 6926430 · Received October 7, 2017

Report

Report Number
1045254-2017-00348
Event Type
Malfunction
Date Received
October 7, 2017
Date of Event
September 11, 2017
Report Date
November 28, 2017
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
UDI-DI
00613994933713
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCT(S): A 1845020 (ATTACHMENT, INDIGO OTOL ANGLED): SERIAL NUMBER - (B)(4); LOT NUMBER ¿ 210461789; MANUFACTURED DATE ¿ DECEMBER 5, 2015; UDI # - (B)(4); 510K - K081475. INITIAL REPORTER'S PHONE NUMBER WITH EXTENSION: (B)(6). A 1845000 (DRILL, INDIGO HIGH SPEED OTOLOGIC): THE DEVICE HAS BEEN RETURNED. THE PRODUCT ANALYSIS RESULTS ARE PENDING COMPLETION OF THE EVALUATION. 1845020 (ATTACHMENT, INDIGO OTOL ANGLED): THE DEVICE HAS NOT BEEN RETURNED. A PRODUCT ANALYSIS HAS NOT BEEN PERFORMED. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

A 1845000, INDIGO DRILL (SERIAL NUMBER (B)(4)): THERE WAS NO OVERHEATING FOUND. THE DRILL WAS RUN FOR 5 MINUTES AT 52,000 RPM AND THE TEMPERATURE WAS 88 DEGREES F. 1845020, INDIGO ANGLED ATTACHMENT (SERIAL NUMBER (B)(4)): THERE WAS NO FAILURE FOUND. THE ANGLED ATTACHMENT WAS RUN FOR 5 MINUTES AT 52,000 RPM AND THE TEMPERATURE WAS 90 DEGREES F. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE ANGLED DRILL HEATED UP WITHIN SECONDS. A REPLACEMENT DRILL WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT IMPACT OR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
705679 IPC® HANDPIECE - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000 210276789 00613994933713

Patients

Seq Age Sex Outcome Treatment
1