FDA Adverse Event Injury Summary report: N

AQUAMANTYS GENERATOR (115V)

MDR report key: 6925803 · Received October 6, 2017

Report

Report Number
1226420-2017-00247
Event Type
Injury
Date Received
October 6, 2017
Date of Event
September 7, 2017
Report Date
October 4, 2018
Manufacturer
MEDTRONIC ADVANCED ENERGY, LLC
Product Code
GEI
PMA / PMN Number
K052859
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BRIEF DESCRIPTION OF COMPLAINT: A 6.0 DEVICE WAS BEING USED TO PROACTIVELY TREAT THE PRESCRIBED AREAS AND CONTINUOUS SUCTION WAS BEING USED; HOWEVER, SOME OF THE HEATED SALINE WAS ABLE TO GET PAST THE SUCTION. THE SALINE RAN DOWN THE PATIENT¿S LEG, RESULTING IN A BURN TO THE UPPER TWO THIRDS OF THE RIGHT MOST ANTERIOR SHIN. INVESTIGATION CONCLUSION: THE REPORTED ISSUE WAS NOT CONFIRMED. GENERATOR PASSES ALL FUNCTIONAL, PRIME, SALINE FLOW, AND POWER OUTPUT TESTS. THERE ARE NO FAILURES FOUND AND THE PATIENT INJURY ALLEGATION CANNOT BE REPLICATED IN THE LABORATORY ENVIRONMENT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

PRODUCT EVENT: (B)(4). PATIENT INFORMATION INCOMPLETE AND MISSING PATIENT INFORMATION UNABLE TO BE OBTAINED DESPITE A GOOD FAITH EFFORT MADE TO OBTAIN THE INFORMATION FROM THE CUSTOMER.

Description of Event or Problem · 1

DURING AN ORTHO CASE, WHILE USING SUCTION, HEATED SALINE FROM THE AQUAMANTYS DEVICE RAN OFF ONTO THE PATIENT¿S LEG RESULTING IN A BURN. THE BURN WAS APPROXIMATELY 4 INCHES LONG, THE DEGREE IS UNKNOWN. THE BURN WAS TREATED WITH AN ANTIMICROBIAL DRESSING.

Description of Event or Problem · 1

DURING AN ORTHO CASE, WHILE USING SUCTION, HEATED SALINE FROM THE AQUAMANTYS DEVICE RAN OFF ONTO THE PATIENT¿S LEG RESULTING IN A BURN. THE BURN WAS APPROXIMATELY 4 INCHES LONG, THE DEGREE IS UNKNOWN. THE BURN WAS TREATED WITH AN ANTIMICROBIAL DRESSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
700489 AQUAMANTYS GENERATOR (115V) ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI MEDTRONIC ADVANCED ENERGY, LLC 40-402-1

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention