MILLENIUM VG DUAL COLLIMATOR

MDR report key: 692347 · Received March 24, 2006

Report

Report Number
9613299-2006-00005
Event Type
Malfunction
Date Received
March 24, 2006
Date of Event
February 22, 2006
Report Date
March 24, 2006
Manufacturer
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Product Code
IYX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT'S WEIGHT IS UNK.

Description of Event or Problem · 1

A VG GAMMA CAMERA COLLIMATOR (VPC-45) PARTIALLY RELEASED FROM THE DETECTOR HEAD. THE USER COMPLETED THREE "LIVER" SCAN PROTOCOLS ON THE PT USING THREE ANGLES. IN ORDER TO PERFORM AN ADDITIONAL SPECT SCAN, THE USER BEGAN TO ROTATE AND ALIGN THE DETECTOR HEADS IN THE 0 AND 180 DEGREE POSITIONS, AND NOTICED THAT THE COLLIMATOR WAS PARTIALLY DETECHED FROM THE FRONT TWO LOCKING MECHANISMS. THE USER ROTATED THE GANTRY TO A SAFE POSITION AND RETRACTED THE PT TABLE OUT FROM UNDER DETECTORS. NO INJURIES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MILLENIUM VG DUAL COLLIMATOR NUCLEAR MEDICINE IYX GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING NI NA

Patients

Seq Age Sex Outcome Treatment
1 72 YR