MULTISTIX 8 SG REAGENT STRIPS FOR URINALYSIS
Report
- Report Number
- 3002637618-2017-00135
- Event Type
- Malfunction
- Date Received
- October 6, 2017
- Date of Event
- August 25, 2017
- Report Date
- February 27, 2018
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS
- Product Code
- JIR
- PMA / PMN Number
- K992257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
SIEMENS TESTED THE SUBMITTED REAGENT STRIPS WHICH SHOWED NO OBVIOUS VISIBLE DAMAGE OR DEGRADATION. THE BOTTLE ARRIVED TIGHTLY CAPPED AND CONTAINING THE EXPECTED DESICCANT. NO PATIENT SAMPLES WERE SUBMITTED. ALL VALUES FOR BOTH TEST SOLUTIONS REPORTED IN THE EXPECTED RANGES: STRIPS TESTED WITH POSITIVE (42 CELL/MCL) LEUKOCYTE SOLUTION DEVELOPED THE CHARACTERISTIC PURPLE COLOR ON THE LEUK PAD, MATCHING THE 30-47 NUMERIC DESIGNATION RANGE FOR SMALL POSITIVE. STRIPS TESTED WITH POSITIVE (0.15 MG/DL) NITRITE SOLUTION DEVELOPED DEFINITE POSITIVE COLOR ON THE NIT PAD, AVERAGING WELL INTO THE 15-30 NUMERIC DESIGNATION RANGE FOR POSITIVE RANGE. CONCLUSION: THE CUSTOMER REPORT OF FALSE LEUKOCYTE AND BACTERIA RESULTS WHEN VISUAL TESTING WITH MULTISTIX 8SG WAS NOT OBSERVED.
THE REAGENT HAS BEEN REQUESTED TO BE RETURNED FOR INVESTIGATION. THE CAUSE OF THIS EVENT IS UNKNOWN.
THE CUSTOMER STATED THAT NO REAGENT WILL BE RETURNED. THE CAUSE OF THE DISCREPANCY IN RESULTS CANNOT BE DETERMINED BASED UPON THE INFORMATION THAT THE CUSTOMER PROVIDED. PROPER TECHNIQUE AND MAINTENANCE WERE REVIEWED WITH THE CUSTOMER. THE CUSTOMER STATED THAT THEY ARE NO LONGER USING MULTISTIX 8 SG STRIPS.
THE CUSTOMER REPORTED A FALSE NEGATIVE LEUKOCYTE AND BACTERIA ON THE MULTISTIX 8 SG REAGENT STRIPS FOR URINALYSIS, READ VISUALLY, WHEN COMPARED TO THE MICROSCOPIC EXAM OF THE SEDIMENT. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 702461 | MULTISTIX 8 SG REAGENT STRIPS FOR URINALYSIS | MULTISTIX 8 SG | JIR | SIEMENS HEALTHCARE DIAGNOSTICS | 702038 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |