QUICKIE QRI
Report
- Report Number
- 9616084-2017-00008
- Event Type
- Malfunction
- Date Received
- October 5, 2017
- Date of Event
- September 7, 2017
- Report Date
- October 24, 2017
- Manufacturer
- SUNRISE MEDICAL PRIVADA MISIONES
- Product Code
- IOR
- PMA / PMN Number
- K123975
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
SUNRISE MEDICAL RECEIVED PART ON (B)(6) 2017 AND AN EVALUATION WAS PERFORMED BY THE QUALITY ANALYST. AFTER INSPECTING THE PART IT WAS DETERMINED THAT SAID PART MEETS SPECIFICATION. THE CAMBER TUBE AS RECEIVED SHOWS NO SIGNS OF FAILURE. THE MOST LIKELY CAUSE OF THE INCIDENT WOULD BE ATTRIBUTED TO A LACK OF OR IMPROPER MAINTENANCE.
SUNRISE MEDICAL RESEARCHED THE HISTORY OF THIS CHAIR. THE ONLY REPAIR THAT WAS RECORDED DONE TO THIS CHAIR WAS THE REPLACEMENT OF A SPACER AND A SET SCREW TO THE QUICK RELEASE AXLE IN (B)(6) OF 2017 BECAUSE THOSE PARTS WERE MISSING. THIS INCIDENT REPORT WAS CONCLUDED WITH A PARTS ORDER FOR THE CAMBER KIT THAT RECEIVES THE QUICK RELEASE AXLE. THIS ITEM WAS SHIPPED TO THE DEALER ON 9/8/2017. SUNRISE MEDICAL HAS REQUESTED FROM THE DEALER THE RETURN OF THESE PARTS FOR AN EVALUATION. THE PARTS HAVE NOT YET BEEN RECEIVED. THERE IS NO INDICATION THAT A MALFUNCTION HAS OCCURRED OR IF THE PARTS ARE DEFECTIVE. IF AND WHEN THE ALLEGED FAILED PART IS RECEIVED AN EVALUATION WILL BE PERFORMED AND A SUPPLEMENTAL REPORT WILL BE FILED WITH ANY RELEVANT FINDINGS.
PLEASE SEE INITIAL MDR 9616084-2017-00008.
PER DEALER (B)(6), THE LEFT WHEEL FELL OFF THE CHAIR WHILE THE CLIENT WAS IN IT. THIS OCCURRED WHILE THE CLIENT WAS IN THE FRONT YARD. THE CLIENT FELL OUT OF THE CHAIR, HOWEVER, NO INJURIES OCCURRED. HE DID NOT SEEK MEDICAL ATTENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 698642 | QUICKIE QRI | WHEELCHAIR, MANUAL | IOR | SUNRISE MEDICAL PRIVADA MISIONES | EIR5 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |