FDA Adverse Event Malfunction Summary report: N

SUTURE PASSER NEEDLE

MDR report key: 6916088 · Received October 5, 2017

Report

Report Number
3004086872-2017-00007
Event Type
Malfunction
Date Received
October 5, 2017
Date of Event
September 6, 2017
Report Date
October 3, 2017
Manufacturer
COORSTEK MEDICAL
Product Code
GAB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE HISTORY RECORD REVIEW: DEVICE HISTORY RECORD REVIEW SHOW NO NON-CONFORMANCES WITH THIS MANUFACTURE LOT. FATIGUE TESTING OF NEEDLES FROM THIS MANUFACTURE LOT SHOWS WITH 95% CONFIDENCE THAT 99.9% OF THE LOT WOULD PASS A MINIMUM OF 24 CYCLES. THE INSTRUCTIONS FOR USE INCLUDED WITH THE PRODUCT CAUTION AGAINST THE USE OF DIRECT FORCE AND INSTRUCT THE USER TO USE THE PRODUCT UNDER DIRECT VISUALIZATION. DATE PRODUCT RECEIVED: PRODUCT WAS NOT RETURNED. (B)(4). VISUAL INSPECTION: WHERE PRODUCT WAS NOT RETURNED A VISUAL INSPECTION WAS NOT PERFORMED. FUNCTIONAL INSPECTION: WHERE PRODUCT WAS NOT RETURNED A FUNCTIONAL INSPECTION WAS NOT PERFORMED. THIS IS A KNOWN FAILURE MODE ADDRESSED IN THE RISK MANAGEMENT FILE FOR THIS PRODUCT, AND DOES NOT EXCEED EXPECTED OCCURRENCE RATES. OCCURRENCE RATES ARE MONITORED FOR TRENDS. SERVICE HISTORY REVIEW: THERE HAS BEEN NO SERVICE ON THIS DEVICE. ROOT CAUSE(S): WHERE THE PRODUCT WAS NOT RETURNED AN EXACT ROOT CAUSE COULD NOT BE DETERMINED. POSSIBLE CAUSES COULD BE USER RELATED SUCH AS THE NEEDLE CONTACTING SOMETHING HARD LIKE BONE. MANUFACTURER ERROR: WITH LIMITED INFORMATION, THE DEVICE HISTORY RECORD WAS REVIEWED AND THERE IS NOTHING IN THE RECORD THAT WOULD ATTRIBUTE THIS TO AN ERROR IN MANUFACTURING.

Description of Event or Problem · 1

CHAMPION SUTURE PASSER NEEDLE BROKE DURING SURGEON ATTEMPT TO PASS SUTURE THROUGH ROTATOR CUFF TENDON. COUNTRY OF EVENT: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
697408 SUTURE PASSER NEEDLE SUTURE PASSER NEEDLE GAB COORSTEK MEDICAL 3910-900-091 1203063

Patients

Seq Age Sex Outcome Treatment
1