FDA Adverse Event Malfunction Summary report: N

NITINOL TC ELECTRODE

MDR report key: 6915094 · Received October 5, 2017

Report

Report Number
3006630150-2017-03385
Event Type
Malfunction
Date Received
October 5, 2017
Date of Event
September 14, 2017
Report Date
September 14, 2017
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250013054
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

MODEL #: TCN-10 LOT #: 102815 DESCRIPTION: NITINOL TC ELECTRODE, 100 MM TOTAL QUANTITY: 1.

Description of Event or Problem · 1

DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES FOUND THAT BOTH DEVICES HAD EPOXY CHIPPED OUT AND DISCOLORATION. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
696097 NITINOL TC ELECTRODE ELECTRODES GXI COSMAN MEDICAL TCN-10 110916 00813250013054

Patients

Seq Age Sex Outcome Treatment
1