FDA Adverse Event
Malfunction
Summary report: N
NITINOL TC ELECTRODE
MDR report key: 6915094
·
Received October 5, 2017
Report
- Report Number
- 3006630150-2017-03385
- Event Type
- Malfunction
- Date Received
- October 5, 2017
- Date of Event
- September 14, 2017
- Report Date
- September 14, 2017
- Manufacturer
- COSMAN MEDICAL
- Product Code
- GXI
- UDI-DI
- 00813250013054
- PMA / PMN Number
- K050084
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
MODEL #: TCN-10 LOT #: 102815 DESCRIPTION: NITINOL TC ELECTRODE, 100 MM TOTAL QUANTITY: 1.
Description of Event or Problem · 1
DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES FOUND THAT BOTH DEVICES HAD EPOXY CHIPPED OUT AND DISCOLORATION. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 696097 | NITINOL TC ELECTRODE | ELECTRODES | GXI | COSMAN MEDICAL | TCN-10 | 110916 | 00813250013054 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |