SYSMEX CA-620
Report
- Report Number
- 1000515253-2017-00034
- Event Type
- Injury
- Date Received
- October 3, 2017
- Date of Event
- August 9, 2017
- Report Date
- October 3, 2017
- Manufacturer
- SYSMEX RA CO. LTD.
- Product Code
- GKP
- PMA / PMN Number
- K031377
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE SAMPLE TUBE (BD 1.8ML DOUBLE WALLED TUBE) USED FOR ANALYSIS WAS UNACCEPTABLE FOR USE ON THE ANALYZER. THE SAMPLE PROBE WAS HITTING THE SIDE OF THE TUBE WHILE ASPIRATING. NO ANALYZER DEFICIENCIES WERE IDENTIFIED. PER THE INSTRUCTIONS FOR USE (IFU) FOR THE ANALYZER, THE ONLY DOUBLE WALLED TUBES TO BE USED ON THE ANALYZER ARE THE VACUTAINER PLUS PLASTIC CITRATE TUBE, 2.7 ML OR THE VACUETTE SANDWICH COAGULATION TUBE, 3.0 ML/3.5ML. THE IFU SPECIFICALLY IDENTIFIES THE TUBE USED AS NOT BEING ACCEPTABLE FOR USE ON THE ANALYZER.
A SAMPLE WAS COLLECTED FROM A PATIENT RECEIVING COUMADIN TREATMENT IN AN UNAPPROVED SAMPLE COLLECTION TUBE. THE USER PERFORMED ANALYSIS WITH THE UNAPPROVED TUBE. THE INITIAL PROTHROMBIN TIME (PT) RESULT WAS "NO COAG", MEANING NO CLOT WAS FORMED AND NO RESULTS COULD BE GENERATED. THE NEXT ANALYSIS WAS A PROTHROMBIN TIME EXTENDED (PTX) WHICH ALSO RESULTED AS "NO COAG". THE PTX WAS REPEATED AND GAVE A RESULT OF 159.4 SECONDS. THIS RESULT WAS REPORTED TO THE CLINICIAN. THE CLINICIAN ADMINISTERED VITAMIN K TO THE PATIENT. THE ENTIRE CLINICAL PICTURE FOR THE PATIENT IS UNKNOWN. THE PATIENT WAS REDRAWN SEVERAL DAYS LATER AND RETURNED A PT RESULT OF 25.6 SECONDS WHICH CORRELATES WITH THE PATIENT'S CLINICAL PICTURE. THE USER REPORTED THIS VALUE IN A CORRECTED REPORT OF THE INITIAL VALUE. THE PATIENT DID NOT SUFFER ANY HARM DUE TO THE VITAMIN K.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 690773 | SYSMEX CA-620 | AUTOMATED BLOOD COAGULATION ANALYZER | GKP | SYSMEX RA CO. LTD. | CA-620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |