FDA Adverse Event
Malfunction
Summary report: N
TABLE, OPERATING-ROOM, ELECTRICAL
MDR report key: 6910954
·
Received October 3, 2017
Report
- Report Number
- 6910954
- Event Type
- Malfunction
- Date Received
- October 3, 2017
- Date of Event
- June 8, 2017
- Report Date
- August 4, 2017
- Manufacturer
- SKYTRON, LLC
- Product Code
- FQO
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
OPERATING ROOM TABLE WAS PUT INTO STEEP TRENDELENBURG POSITION, CAUSING SAFETY BRACKETS ON BED TO DISENGAGE PART OF THE TABLE. PATIENT SLID TOWARD ANESTHESIA ABOUT 4-5 INCHES, WAS IMMEDIATELY NOTICED AND CAUGHT/SUPPORTED BY OPERATING ROOM STAFF AND SAFELY REPOSITIONED ON BED. NO INJURY WAS NOTED AT THIS TIME. CASE WAS CONVERTED TO OPEN DUE TO INABILITY TO POSITION PATIENT SAFELY IN TRENDELENBURG. PATIENT WAS REPREPPED, REDRAPED, AND CASE CONTINUED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692054 | TABLE, OPERATING-ROOM, ELECTRICAL | FQO | SKYTRON, LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |