FDA Adverse Event Malfunction Summary report: N

PICCOLO COMPOSITE NAILING SYSTEM

MDR report key: 6910021 · Received October 3, 2017

Report

Report Number
9615128-2017-00046
Event Type
Malfunction
Date Received
October 3, 2017
Date of Event
September 6, 2017
Report Date
October 3, 2017
Manufacturer
CARBOFIX ORTHOPEDICS LTD.
Product Code
HSB
UDI-DI
07290109335469
PMA / PMN Number
K153536
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EXAMINATION OF THE PRODUCTION RECORDS OF THE INVOLVED DEVICE DID NOT REVEAL A DEVIATION FROM SPECIFICATION THAT MAY HAVE CONTRIBUTED TO THE REPORTED CASE. BIOMECHANICAL TESTS HAVE BEEN PERFORMED TO SIMILAR PRODUCTS. ALL TESTED PRODUCTS SUCCESSFULLY PASSED THE BIOMECHANICAL REQUIREMENTS. IT WAS EMPHASIZE TO THE SURGEON, THE IMPORTANCE OF DRILLING AND TAPPING IN ACCORDANCE WITH THE LAG SCREW LENGTH, AS INSTRUCTED IN THE IFU "PICCOLO COMPOSITE® NAILING SYSTEM - PROXIMAL FEMUR INSTRUCTION FOR USE" 13161_014ROW - REV 08/17.

Description of Event or Problem · 1

DURING IMPLANTATION OF A PICCOLO COMPOSITE PROXIMAL FEMUR NAIL IN (B)(6), A LAG SCREW BROKE. IN THE TIME OF THE SCREW INSERTION, THE SURGEON FELT RESISTANCE AND DECIDED TO EXTRACT THE SCREW IN ORDER TO REPEAT THE DRILLING. DURING UNSCREWING, THE DISTAL TIP OF THE SCREW BROKE AND THE SURGEON EXTRACTED THE SCREW LEAVING THE DISTAL END IN SITU. ANOTHER CARBOFIX LAG SCREW WAS INSERTED AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690367 PICCOLO COMPOSITE NAILING SYSTEM PROXIMAL FEMUR LAG SCREW HSB CARBOFIX ORTHOPEDICS LTD. PPF01109 07290109335469

Patients

Seq Age Sex Outcome Treatment
1 46 YR