FDA Adverse Event Malfunction Summary report: N

IPC® HANDPIECE - INDIGO¿ DRILL

MDR report key: 6909435 · Received October 3, 2017

Report

Report Number
1045254-2017-00342
Event Type
Malfunction
Date Received
October 3, 2017
Date of Event
September 6, 2017
Report Date
March 8, 2018
Manufacturer
MEDTRONIC XOMED INC.
Product Code
HBE
UDI-DI
00613994933713
PMA / PMN Number
K081475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

WE REPORTED 3 DRILLS AND 2 ATTACHMENTS ON MDR 1045254-2017-00342. THIS SUPPLEMENTAL IS TO PROVIDE INFORMATION FOR 1 DRILL, AND THE 2 ATTACHMENTS. INFORMATION ON THE OTHER DRILLS WILL BE PROVIDED IN SUBSEQUENT MDRS, ONE REPORT FOR EACH DRILL. A TECHNICIAN PERFORMED ON SITE TESTING, THIS DEVICE PASSED AND WAS NOT RETURNED TO THE MANUFACTURE FOR ANALYSIS. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

THERE WERE A TOTAL OF 3 HANDPIECES WHERE ISSUES WERE REPORTED, NOT EIGHT. THE ORIGINAL MDR FILE INDICATED THERE WERE "FIVE HANDPIECES OF PRODUCT #1845000 - HANDPIECE INDIGO DRILL, SERIAL AND LOT NUMBER UNKNOWN", THIS INFORMATION WAS INCORRECT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: 1845010 - ATTACHMENT INDIGO STRAIGHT, SERIAL # (B)(4), LOT # 207374998, MANUFACTURED DATE: 09/03/2013, 510(K) # K081475, UDI (B)(4); 1845010 - ATTACHMENT INDIGO STRAIGHT, SERIAL # (B)(4), LOT # 213100831, MANUFACTURED DATE: 04/12/2017, 510(K) # K081475, UDI (B)(4); 1845000 - HANDPIECE INDIGO DRILL, SERIAL # (B)(4), LOT # 212934866, MANUFACTURED DATE: 03/09/2017, 510(K) # K081475, UDI (B)(4); 1845000 - HANDPIECE INDIGO DRILL, SERIAL # (B)(4), LOT # 207089289, MANUFACTURED DATE: 06/17/2013, 510(K) # K081475, UDI: (B)(4); FIVE HANDPIECES OF PRODUCT # 1845000 - HANDPIECE INDIGO DRILL, SERIAL AND LOT NUMBER UNKNOWN. NO ANALYSIS RESULTS AVAILABLE. DEVICE WERE NOT RETURNED.

Description of Event or Problem · 1

THE FIELD TERRITORY MANAGER REPORTED THAT DURING DEVICE FOLLOW-UP, HE WENT THROUGH CHECKING 8 MOTORS WITH TWO STRAIGHT ATTACHMENTS. ONCE THE HANDPIECES REACHED 50K SPEED THEY UNLOCKED AND CAUSED THE MOTOR TO HEAT UP. HE TRIED EACH ATTACHMENT ON EACH MOTOR AND ENDED UP WITH THE SAME RESULT. THERE IS NO PATIENT INVOLVED IN THIS EVENT, NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692033 IPC® HANDPIECE - INDIGO¿ DRILL DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE MEDTRONIC XOMED INC. 1845000 206315728 00613994933713

Patients

Seq Age Sex Outcome Treatment
1